The size requirements in this section apply only to the portion of the diagnostic equipment used for transfer....
Search Results "Examination Equipment"
-
M301.2.2 Size
-
M302.4.2 Clearance Around Base
The base of the equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination surface...
-
4. Exception from the Requirements of M301 for Certain Examination Chairs that Comply with M302
The Access Board proposed in M101.2 in the MDE NPRM to require diagnostic equipment to meet the standards for each patient position supported, meaning that if diagnostic equipment was designed...
-
M302.4.2 Clearance Around Base
The base of the diagnostic equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination...
-
5.4.10 Transfer Support Fittings for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.10 Transfer Support Fittings for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
M305.3.2 Height
The height of gripping surfaces for standing supports shall be based on the position of the standing supports in relation to the standing surfaces they serve. Horizontal standing...
-
M305.3.2 Height (Section-by-Section Analysis)
For horizontal supports, the MDE NPRM proposed a gripping surface height of 34 inches minimum and 38 inches maximum above the standing surface. There were no public comments on this...
-
7. Diagnostic Imaging Equipment: Accessibility Considerations
and transfer supports, for current diagnostic imaging equipment....
-
M301.3.3 Head and Back Support
M301.3.3 Head and Back Support Proposed requirement: Where diagnostic equipment used by patients in a supine, prone, or side-lying position, and diagnostic equipment used by patients...
-
M305.3.1.1 Horizontal Position
The length of gripping surfaces on horizontal standing supports shall be 4 inches (100 mm) minimum except for diagnostic equipment containing a wheelchair space which shall comply with M305.3.1.2...
-
A. Final Regulatory Assessment (EO 13563 and EO 12866)
Examples of such diagnostic equipment include examination tables and chairs, weight scales, mammography equipment, and other imaging equipment....
-
5.4.5 Transfer Support Position for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.5 Transfer Support Position for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces. Standing supports should be located on both sides of platforms, a minimum of 34 inches...
-
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations The Committee recommends standing supports located on one side of the...
-
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations The Committee recommends standing supports located on both sides of...
-
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces. Standing supports should be located on one side of the platform, integrated into the platform...
-
5.7.1 Lift Compatibility Clearance in Base for Equipment Used by Patient in Supine, Prone, or Side-Lying Position
5.7.1 Lift Compatibility Clearance in Base for Equipment Used by Patient in Supine, Prone, or Side-Lying Position....
-
5.4.1 Transfer Support Location for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.1 Transfer Support Location for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
5.4.2 Transfer Support Length for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.2 Transfer Support Length for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
Figure M5: Overview of M303 Diagnostic Equipment for Use by a Patient While Seated in a Wheelchair and 304 Diagnostic Equipment for Use by a Patient in a Standing Position (Mammography Machine Example)
Where the primary use of the equipment supports more than one patient position, it must comply with the requirements for each use. ...
-
M301.3.2 Leg Supports (Section-by-Section Analysis)
In the MDE NPRM, the Access Board proposed to place the requirements for stirrups on diagnostic equipment used by patients in the supine, prone, or side-lying position in M301....
-
M301.4 Lift Compatibility
Where the width of the examination surface is less than 36 inches (915 mm), the clearance depth shall extend the full width of the equipment....
-
M301.4.2 Clearance Around Base
[See subsections ...]
-
5.4.9 Interruptions Along Transfer Support Gripping for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.9 Interruptions Along Transfer Support Gripping for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....