Transfer supports shall have a cross section complying with 305.2.4.1 or 305.2.4.2.
Search Results "Examination Equipment"
-
M305.2.4 Cross Section
-
M305.2.4.1 Circular Cross Section
Transfer supports with circular cross sections shall have an outside diameter of 1 ¼ inches (32 mm) minimum and 2 inches (51 mm) maximum.
-
M305.2.4.2 Non-Circular Cross Section
Transfer supports with non-circular cross sections shall have a cross-section dimension of 2 inches (51 mm) maximum and a perimeter dimension of 4 inches (100 mm) minimum and 4.8...
-
M305.2.5 Surface Hazards
Transfer supports and surfaces adjacent to transfer supports shall be free of sharp or abrasive components and shall have eased edges.
-
M305.2.6 Gripping Surface
Transfer support gripping surfaces shall be continuous along their length and shall not be obstructed along their tops or sides. The bottoms of transfer support gripping surfaces...
-
M305.2.7 Clearance
Clearance between the transfer support gripping surface and adjacent surfaces or obstructions shall be 1 ½ inches (38 mm) minimum.
-
M305.2.8 Fittings
Transfer supports shall not rotate within their fittings when in place for transfer.
-
M305.3 Standing Supports
Standing supports shall provide continuous support throughout use of the diagnostic equipment and shall comply with M305.3....
-
M305.3 Standing Supports
Standing supports shall provide continuous support throughout use of the diagnostic equipment and shall comply with M305.3....
-
M305.3.1 Length
Horizontal standing support gripping surfaces shall comply with M305.3.1.1, horizontal standing support gripping surfaces on diagnostic equipment containing a wheelchair space shall comply...
-
M305.3.1 Horizontal Position
Where the support is horizontal, the top of the gripping surface shall be 34 inches (865 mm) minimum and 38 inches (965 mm) maximum above the standing surface. The gripping surface...
-
M305.3.1.1 Horizontal Position
The length of gripping surfaces on horizontal standing supports shall be 4 inches (100 mm) minimum except for diagnostic equipment containing a wheelchair space which shall comply with M305.3.1.2...
-
M305.3.1.2 Diagnostic Equipment Containing a Wheelchair Space
On diagnostic equipment containing wheelchair spaces with one entry that also serves as the exit, the length of the gripping surface of horizontal standing supports shall be equal to or...
-
M305.3.1.3 Vertical Position
The length of the gripping surface on vertical standing supports shall be 18 inches (455 mm) minimum.
-
M305.3.2 Vertical Position
Where the support is vertical, it shall be 18 inches (455 mm) minimum in length and the bottom end of the support shall be 34 inches (865 mm) high minimum and 37 inches (940 mm) high...
-
M305.3.2 Height
The height of gripping surfaces for standing supports shall be based on the position of the standing supports in relation to the standing surfaces they serve. Horizontal standing...
-
M305.3.2.1 Horizontal Position
The height of the top of the gripping surface on horizontal standing supports shall be 34 inches (865 mm) minimum and 38 inches (965 mm) maximum above the standing surface.
-
M305.3.2.2 Vertical Position
The height of the lowest end of the gripping surface on vertical standing supports shall be 34 inches (865 mm) minimum and 37 inches (940 mm) maximum above the standing surface.
-
M305.3.3 Fittings
Standing supports shall not rotate within their fittings.
-
M305.3.3 Fittings
Standing supports shall not rotate within their fittings.
-
M305.4 Leg Supports
Leg supports shall provide a method of supporting, positioning, and securing the patient’s legs.
-
M305.5 Head and Back Support
Where the diagnostic equipment is used in a reclined position, head and back support shall be provided....
-
M306 Communication
-
M306 Communication