36 CFR Part 1195 Proposed Accessibility Standards for Medical Diagnostic Equipment - Example Applications of Proposed Standards
-
36 CFR Part 1195 Standards for Accessible Medical Diagnostic Equipment - Preamble
The standards for medical diagnostic equipment (MDE Standards) contain minimum technical criteria to ensure that medical diagnostic equipment, including but not limited to, examination tables...
-
36 CFR Part 1195 Proposed Accessibility Standards for Medical Diagnostic Equipment NPRM
The proposed standards contain minimum technical criteria to ensure that medical diagnostic equipment, including examination tables, examination chairs, weight scales, mammography equipment...
-
36 CFR Part 1195 Standards for Accessible Medical Diagnostic Equipment
The standards for medical diagnostic equipment (MDE Standards) contain minimum technical criteria to ensure that medical diagnostic equipment, including but not limited to, examination tables...
-
36 CFR Part 1195 Proposed Accessibility Standards for Medical Diagnostic Equipment NPRM - Preamble
The proposed standards contain minimum technical criteria to ensure that medical diagnostic equipment, including examination tables, examination chairs, weight scales, mammography equipment...
-
Recommendations on Standards for the Design of Medical Diagnostic Equipment for Adults with Disabilities, Advisory Committee Final Report
Advancing Equal Access to Diagnostic Services: Recommendations on Standards for the Design of Medical Diagnostic Equipment for Adults with Disabilities The final report of the Medical...
-
Figure M1: Overview of M301 Diagnostic Equipment for Use by a Patient in a Supine, Prone, or Side-Lying Position
-
5.1 Transfer Surface Height
5.1 Transfer Surface Height
-
5.1.1 Transfer Surface Height Adjustability
5.1.1 Transfer Surface Height Adjustability. The transfer surface should have an adjustable height in small virtually continuous increments.
-
5.1.2 Transfer Surface High Height
5.1.2 Transfer Surface High Height. The transfer surface should have an adjustable height range with an upper height measuring at least 25 inches.
-
5.1.3 Transfer Surface Low Height
5.1.3 Transfer Surface Low Height. No agreement on final low height
-
5.1.4 Transfer Surface Height Measurement
5.1.4 Transfer Surface Height Measurement. The transfer surface height should be measured from the floor to the highest point on the seat in an uncompressed state, inclusive of...
-
5.2 Transfer Surface Size for Equipment Used by Patients in Supine, Prone, or Side-Lying Position
5.2 Transfer Surface Size for Equipment Used by Patients in Supine, Prone, or Side-Lying Position....
-
5.2.1 Transfer Surface Size
5.2.1 Transfer Surface Size
-
5.2.1.1 Transfer Surface Width
The transfer surface should be 28 inches wide minimum for equipment used by patients in a supine, side-lying, or prone position....
-
5.2.1.2 Transfer Surface Depth
The transfer surface should be 17 inches deep minimum for all equipment used by patients in a supine, side-lying, or prone position with the exception of imaging equipment....
-
5.2.1.3 Transfer Surface Size for Stretchers
5.2.1.3 Transfer Surface Size for Stretchers. The transfer surface should be 28 inches wide minimum and 17 inches deep minimum located on both long sides of stretchers.
-
5.2.1.4 Transfer Surface Size for Imaging Equipment
5.2.1.4 Transfer Surface Size for Imaging Equipment. The transfer surface should be 28 inches wide minimum and 21 inches deep minimum....
-
5.2.2 Transfer Surface Size for Equipment Used by Patients in Seated Position
5.2.2 Transfer Surface Size for Equipment Used by Patients in Seated Position....
-
5.2.2.1 Transfer Surface Width
The transfer surface should be 21 inches wide minimum for examination chairs and other equipment used by patients in a seated position....
-
5.2.2.2 Transfer Surface Depth
The transfer surface should be 17 inches deep minimum for equipment used by patients in a seated position....
-
5.2.3 Transfer Surface Measurement
5.2.3 Transfer Surface Measurement. The transfer surface should be measured from the center-point of each side of the transfer surface.
-
5.3 Transfer Sides
5.3 Transfer Sides.
-
5.3.1 Permitted Obstructions to Transfer Sides for Equipment Used by Patients in a Supine, Prone, Side-Lying or Seated Position
5.3.1 Permitted Obstructions to Transfer Sides for Equipment Used by Patients in a Supine, Prone, Side-Lying or Seated Position....