A transfer surface shall be provided and shall comply with M301.2.
Search Results "Examination Equipment"
-
M301.2 Transfer Surface
-
M301.2 Transfer Surface
A transfer surface shall be provided and shall comply with M301.2.
-
M301.2.1 Adjustability
Transfer surfaces shall be adjustable in height measured from the floor to the top of the uncompressed transfer surface and shall provide the following: A. A low transfer position...
-
M301.2.1 Height
The height of the transfer surface during patient transfer shall be 17 inches (430 mm) minimum and 19 inches (485 mm) maximum measured from the floor to the top of the transfer...
-
M301.2.2 Sunset
The low transfer position height, Item A of M301.2.1, shall cease to have effect on [Insert date 5 years after date of publication in the Federal Register].
-
M301.2.2 Size
The size requirements in this section apply only to the portion of the diagnostic equipment used for transfer....
-
M301.2.3 Transfer Sides
The transfer surface shall be located to provide options to transfer from a mobility device onto one short side (depth) and one long side (width) of the surface. Each transfer side...
-
M301.2.3 Size
The size of the transfer surface shall comply with M301.2.3.1 or M301.2.3.2. The size of transfer surfaces shall be measured from center points of their opposing sides.
-
M301.2.3.1 End Transfer Surface
EXCEPTION: Transfer surfaces for imaging equipment with bores shall be permitted to be 21 inches (535 mm) wide minimum but shall not be permitted to be less than the full width of the examination...
-
M301.2.3.2 Side Transfer Surface
EXCEPTION: Transfer surfaces for imaging equipment with bores shall be permitted to be 21 inches (535 mm) wide minimum but shall not be permitted to be less than the full width of the examination...
-
M301.2.4 Unobstructed Transfer
Each transfer surface shall provide two unobstructed sides for patient transfer. EXCEPTIONS: 1. Obstructions no more than 3 inches (75 mm) deep shall be permitted to extend beyond...
-
M301.3 Supports
Transfer supports, leg supports, and reclining surfaces shall comply with M301.3.
-
M301.3 Supports
Transfer supports, stirrups, and reclining surfaces shall comply with M301.3.
-
M301.3.1 Transfer Supports
Transfer supports shall be provided for use with the transfer sides required by M301.2.3 and shall comply with M305.2.
-
M301.3.1 Transfer Supports
Transfer surfaces required by M301.2 shall provide transfer supports and shall comply with M305.2.
-
M301.3.2 Leg Supports
Where stirrups are provided, leg supports shall also be provided and shall comply with M305.4.
-
M301.3.2 Stirrups
Where stirrups are provided, they shall provide a method of supporting, positioning, and securing the patient’s legs.
-
M301.3.3 Head and Back Support
Where the diagnostic equipment is used in a reclined position, head and back support shall be provided....
-
M301.3.3 Head and Back Support
Where the diagnostic equipment is used in a reclined position, head and back support shall be provided and shall comply with M305.5....
-
M301.4 Lift Compatibility
Diagnostic equipment shall be usable with a patient lift and shall comply with M301.4.1 or M301.4.2....
-
M301.4 Lift Compatibility
Diagnostic equipment shall be usable with portable patient lifts and, when in use with such lifts, shall comply with M301.4.1 or M301.4.2....
-
M301.4.1 Clearance in Base
from the edge of the examination surface....
-
M301.4.1 Clearance in Base
from the edge of the examination surface....
-
M301.4.2 Clearance Around Base
The base of diagnostic equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination...