Hello. Please sign in!

36 CFR Part 1195 Proposed Accessibility Standards for Medical Diagnostic Equipment NPRM

This is an NPRM for the Proposed MDE Standards (2012). Click here to view the Final MDE Standards (2017).

M305.3 Standing Supports.

Standing supports shall provide continuous support throughout use of the diagnostic equipment and shall comply with M305.3.

M305.3.1 Horizontal Position.

Where the support is horizontal, the top of the gripping surface shall be 34 inches (865 mm) minimum and 38 inches (965 mm) maximum above the standing surface. The gripping surface shall be 4 inches (100 mm) long minimum.

M305.3.2 Vertical Position.

Where the support is vertical, it shall be 18 inches (455 mm) minimum in length and the bottom end of the support shall be 34 inches (865 mm) high minimum and 37 inches (940 mm) high maximum above the standing surface.

M305.3.3 Fittings.

Standing supports shall not rotate within their fittings.

[MORE INFO...]

*You must sign in to view [MORE INFO...]