5.7.1.1 Lift Compatibility Clearance in Base for Stretchers. The base of the equipment must provide a clearance of 39 inches wide minimum.
Search Results "Examination"
Commonly Searched Documents
-
5.7.1.1 Lift Compatibility Clearance in Base for Stretchers
-
5.7.1.1 Lift Compatibility Clearance in Base for Stretchers (M301)
5.7.1.1 Lift Compatibility Clearance in Base for Stretchers (M301) The Committee recommends the equipment base provide a clearance of 39 inches wide minimum. Rationale for the...
-
M301.4 Lift Compatibility
equipment shall provide a clearance 44 inches (1120 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep minimum measured from the edge of the examination...
-
6. Minimum (Low) Height Standard for Transfer Surfaces
However, despite extensive discussion, review of available evidence, and recommendations from lengthy deliberations of the Examination Tables and Chairs Subcommittee (Section 3.2), the Advisory...
-
A. Medical equipment and furniture
Medical equipment and furniture Without accessible medical examination tables, dental chairs, radiological diagnostic equipment, scales, and rehabilitation equipment, individuals with...
-
M303.2.4.1 Toe Clearance
Toe clearance shall be provided at a height of 9 inches (230 mm) above the floor to a depth of 6 inches (150 mm) maximum.
-
M303.2.4.2 Knee Clearance
Knee clearance shall be provided at a depth of 11 inches (280 mm) minimum and 25 inches (635 mm) maximum at 9 inches (230 mm) above the floor and at a depth of 8 inches (205 mm)...
-
M305.3.1.1 Horizontal Position
The length of gripping surfaces on horizontal standing supports shall be 4 inches (100 mm) minimum except for diagnostic equipment containing a wheelchair space which shall comply...
-
M303.2.4.2 Other Equipment
Wheelchair spaces beneath diagnostic equipment other than breast platforms shall comply with M303.2.4.2.
-
11B-805.4.2 Equipment
Clear space complying with Section 11B-305.2 shall be provided as required for specific equipment.
-
M302.2.5 Unobstructed Transfer
Each transfer side complying with M302.2.4 shall provide unobstructed access to transfer surfaces. EXCEPTIONS: 1. Obstructions no more than 3 inches (75 mm) deep shall be...
-
M305.3.2.1 Horizontal Position
The height of the top of the gripping surface on horizontal standing supports shall be 34 inches (865 mm) minimum and 38 inches (965 mm) maximum above the standing surface.
-
M302.2.5 Unobstructed Transfer (Section-by-Section Analysis)
In the MDE NPRM the Access Board proposed that each transfer side provide unobstructed access to the transfer surface, with an exception to permit temporary obstructions as long as...
-
5.6.1 Stirrups for Equipment Used by Patient in Supine, Prone, or Side-Lying Position
5.6.1 Stirrups for Equipment Used by Patient in Supine, Prone, or Side-Lying Position. Where the equipment provides stirrups, it must also provide an alternate method to support,...
-
5.8.1.1 Clear Platform Size Recommendation
5.8.1.1 Clear Platform Size Recommendation The Committee recommends a platform size with a minimum clear width of 32 inches and a minimum clear length of 40 inches. Rationale...
-
5.8.1.1 Clear Platform Size
5.8.1.1 Clear Platform Size. The platform should be a minimum clear width of 32-inches and a minimum clear length of 40-inches.
-
5.6.1 Stirrups Provision for M301
5.6.1 Stirrups Provision for M301 NPRM Description: For patients with limited leg strength and control, thigh, knee and calf supports are necessary to maintain an appropriate...
-
11B-805.4.2 Equipment
Clear space complying with Section 11B-305.2 shall be provided as required for specific equipment.
-
11B-805.4.2 Equipment
Clear space complying with Section 11B-305.2 shall be provided as required for specific equipment.
-
M301.4.1 Clearance in Base
equipment shall provide a clearance 39 inches (990 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep minimum measured from the edge of the examination...
-
M301.4.1 Clearance in Base
equipment shall provide a clearance 44 inches (1120 mm) wide minimum, 6 inches (150 mm) high minimum measured from the floor, and 36 inches (915 mm) deep minimum measured from the edge of the examination...
-
M301.4.2 Clearance Around Base
The base of the equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination surface...
-
M301.4.2 Clearance Around Base
The base of diagnostic equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination...
-
M301.4.2 Clearance Around Base (Section-by-Section Analysis)
The proposed rule required a minimum clearance of 6 inches high measured from the floor and 36 inches deep measured from the edge of the examination surface. NPRM, 77 FR at 6927....