A transfer surface shall be provided and shall comply with M302.2.
Search Results "Examination"
Commonly Searched Documents
-
M302.2 Transfer Surface
-
M302.3.1 Transfer Supports
Transfer supports shall be provided for use with transfer sides required by M302.2.4 and shall comply with M305.2.1.1, M305.2.2.1, and M305.2.3 through M305.2.8.
-
M307 Operable Parts
[See subsections ...]
-
M302.3.1 Transfer Supports (Section-by-Section Analysis)
In the MDE NPRM the Access Board proposed that transfer supports must be provided for use with transfer sides on diagnostic equipment used by patients in the seated position, and...
-
M302.2.3 Size (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface size for diagnostic equipment used by patients in the seated position of 21 inches wide and 15 inches deep (proposed M302.2.2). The Access...
-
M302.2 Transfer Surface (Section-by-Section Analysis)
This is an introductory section.
-
M302.3 Supports (Section-by-Section Analysis)
This is an introductory section. An editorial change was made to this section as a result of the change in M302.3.2, described below, to replace the word “stirrups” with the term...
-
5.2.2.2 Transfer Surface Depth Recommendation for M302
5.2.2.2 Transfer Surface Depth Recommendation for M302 The Committee recommends a minimum depth requirement of 17 inches for equipment used by patients in seated...
-
5.2.2 Transfer Surface Size for Equipment Used by Patients in Seated Position
5.2.2 Transfer Surface Size for Equipment Used by Patients in Seated Position.
-
5.2.2.2 Transfer Surface Depth
5.2.2.2 Transfer Surface Depth. The transfer surface should be 17 inches deep minimum for equipment used by patients in a seated position.
-
M302.2.2 Size
The transfer surface shall be 21 inches (610 mm) wide minimum and 15 inches (381 mm) deep minimum. Advisory M302.2.2 Size. The size requirements in this section apply only to the...
-
M302.3.1 Transfer Supports
Transfer supports shall be provided for use with the transfer sides required by M302.2.3 and shall comply with M305.2.
-
M302.2 Transfer Surface
A transfer surface shall be provided and shall comply with M302.2.
-
M302.3 Supports
Transfer supports, armrests, and reclining surfaces shall comply with M302.3.
-
M307 Operable Parts
[See subsections ...]
-
4. Exception from the Requirements of M301 for Certain Examination Chairs that Comply with M302
The Access Board concurs with commenters that in this limited situation it is unnecessary for the examination chair, which complies with the technical requirements in M302, to also have...
-
Emphasis on Limitations Instead of Outcomes
National Board of Medical Examiners, Gonzales v. National Board of Medical Examiners, and Wong v. Regents of University of California....
-
M302 Diagnostic Equipment used by Patients in a Seated Position (Section-by-Section Analysis)
M302 in the final rule establishes the technical criteria for diagnostic equipment used by patients in a seated position such as examination chairs....
-
PART 4: ACCESSIBLE MEDICAL EQUIPMENT
For example, if a patient must be lying down to be thoroughly examined, then a person with a disability must also be examined lying down....
-
M305.3 Standing Supports
Standing supports shall provide continuous support throughout use of the diagnostic equipment and shall comply with M305.3.
-
M305.3 Standing Supports (Section-by-Section Analysis)
M305.3 provides the technical requirements for standing supports which are required on diagnostic equipment covered by M304. This provision has been reorganized in the final rule...
-
5.9 Standing Support for Equipment Used by Patients in a Standing Position
5.9 Standing Support for Equipment Used by Patients in a Standing Position.
-
5.9 Standing Support Recommendations for M304 Diagnostic Equipment Used by Patients in a Standing Position
5.9 Standing Support Recommendations for M304 Diagnostic Equipment Used by Patients in a Standing Position Description: Many people who ambulate with a disability need support to...
-
M305.3 Standing Supports
Standing supports shall provide continuous support throughout use of the diagnostic equipment and shall comply with M305.3.