M306 in the final rule provides the technical criteria for communication from the diagnostic equipment to the patient.
Search Results "Examination"
Commonly Searched Documents
-
M306 Communication (Section-by-Section Analysis)
-
M302.1 General
Where diagnostic equipment is used by patients in a seated position, it shall comply with M302.
-
M307.2 Tactilely Discernible
Operable parts shall be tactilely discernible without activation.
-
M306 Communication
[See subsections ...]
-
M306.1 General
Where instructions or other information is communicated to the patient through the diagnostic equipment, the instructions and other information shall be provided in at least two of...
-
M307 Operable Parts (Section-by-Section Analysis)
For example, equipment used for an auditory examination may require the patient to press a button when sounds are heard....
-
Section 36.309 Examinations and Courses (Section-by-Section Analysis)
Section 36.309 Examinations and Courses (Section-by-Section Analysis) Section 36.309(a) sets forth the general rule that any private entity that offers examinations or courses relating...
- Midmark Ritter 222 Barrier-Free® Power Examination Table
- Midmark Ritter 223 Barrier-Free® Power Examination Table
-
5.3.3.1 Transfer Sides for Exam Chairs with Fixed Footrests
The 17 inch minimum depth and 21 inch minimum width should be located on both sides of examination chairs with a fixed footrest to allow for a left or right transfer....
-
M302.3.2 Leg Supports (Section-by-Section Analysis)
This will ensure that patients with limited leg strength and control will be able to keep their legs in the appropriate position for examination....
-
11B-805.4.2 Equipment
Clear space complying with Section 11B-305.2 shall be provided as required for specific equipment.
-
Sec.36.309(a) General
Any private entity that offers examinations or courses related to applications, licensing, certification, or credentialing for secondary or postsecondary education, professional, or trade...
-
§ 36.309(a) General
Any private entity that offers examinations or courses related to applications, licensing, certification, or credentialing for secondary or postsecondary education, professional, or trade...
-
Features of an Accessible Examination Room
Features of an Accessible Examination Room 1....
-
5.3.3.1 Transfer Sides Recommendation for Exam Chairs with Fixed Footrests
5.3.3.1 Transfer Sides Recommendation for Exam Chairs with Fixed Footrests Some examination chairs have footrests that get in the way of entry from the foot end, such as dental chairs...
-
ADA Requirements - Testing Accommodations
Department of Justice Civil Rights Division Disability Rights Section Testing Accommodations Standardized examinations and other high-stakes tests are gateways to educational...
-
5.9.1 Standing Supports Recommendation In Reference To Mammography Equipment
5.9.1 Standing Supports Recommendation In Reference To Mammography Equipment The Committee recommends removing the requirement for standing supports on mammography equipment....
-
5.9.1 Standing Supports In Reference To Mammography Equipment
5.9.1 Standing Supports In Reference To Mammography Equipment. The requirement for standing supports should be removed for mammography equipment. To be useful, positioning supports...
-
5.5.1 Armrest Provision for Equipment Used by Patient in Seated Position
5.5.1 Armrest Provision for Equipment Used by Patient in Seated Position. Armrests are not required but if provided cannot obstruct transfer supports.
-
5.3.3 Transfer Sides Recommendations for M302
This provision will coincide with the seat area design of most examination chairs. NPRM Proposed Provision: M302.2.3 Transfer Sides....
-
M302.2.3 Size
Transfer surfaces shall be 21 inches (610 mm) wide minimum and 17 inches (430 mm) deep minimum. The size of transfer surfaces shall be measured from center points of their opposing...
-
M302.3.2 Leg Supports
Where stirrups are provided, leg supports shall also be provided and comply with M305.4.
-
M302.3 Supports
Transfer supports, leg supports and reclining surfaces shall comply with M302.3.