Where wheelchair spaces can be entered from the front or rear, the wheelchair space shall be 48 inches (1220 mm) deep minimum....
Search Results "Wheelchair Space (Medical Diagnostic Equipment)"
Commonly Searched Documents
-
M303.2.3 Depth
-
2. Equipment Clearances for Breast Platforms
The MDE NPRM proposed knee and toe clearance requirements for diagnostic equipment used by patients seated in wheelchairs that paralleled the knee and toe clearance requirements from the...
-
M101.1 Purpose
These Standards (MDE Standards) contain scoping and technical requirements for medical diagnostic equipment (diagnostic equipment) to ensure accessibility to, and usability of the diagnostic...
-
M303.2.4 Equipment Clearances (Section-by-Section Analysis)
The MDE NPRM proposed knee and toe clearance for diagnostic equipment used by patients seated in wheelchairs to allow for components in the wheelchair space which the patient could approach...
-
Equipment and furniture. (Section-by-Section Analysis)
especially for medical equipment....
-
Equipment and furniture. (Section-by-Section Analysis)
Among the commenters' key concerns, many from the disability community objected to the Department's earlier decision not to issue equipment regulations, especially for medical equipment....
-
M101.1 Purpose
The standards contain technical criteria for medical diagnostic equipment that is accessible to and usable by patients with disabilities....
-
M303.2.3 Depth
The depth of wheelchair spaces shall comply with M303.2.3....
-
M303.2.2 Width
Wheelchair spaces shall be 36 inches (915 mm) wide minimum....
-
Knee & Toe Space [M303.2.4]
Each of these measurements creates a more accessible piece of imaging equipment balanced with the physical attributes and operation of the equipment....
-
Wheeled Mobility Anthropometry Project
spaces, including knee and toe clearance, do not accommodate many people in the sample....
-
M303.1 General
Diagnostic equipment used by patients seated in a wheelchair shall comply with M303....
-
M303.1 General
Diagnostic equipment used by patients seated in a wheelchair shall comply with M303....
-
M303.2.2 Width (Section-by-Section Analysis)
The MDE NPRM proposed to require that diagnostic equipment used by patients seated in a wheelchair provide a wheelchair space that was at least 36 inches wide....
-
M101.3 Existing Diagnostic Equipment
The MDE Standards do not address the applicability of scoping or technical requirements to existing diagnostic equipment....
-
M303.4 Components
Where components of diagnostic equipment are used to examine specific body parts, the components shall be capable of examining the body parts of a patient seated in a wheelchair....
-
M303.4 Components
Where components of diagnostic equipment are used to examine specific body parts, the components shall be capable of examining the body parts of a patient seated in a wheelchair....
-
M303.2.4.2.2 Height
Wheelchair spaces shall include toe clearance 9 inches (230 mm) high minimum above the floor measured to a depth of 6 inches (150 mm) maximum measured from the toe end of the wheelchair...
-
H. Barriers Affecting Accessibility and Usability of Medical Diagnostic Equipment
The Rehabilitation Engineering Research Center on Accessible Medical Instrumentation conducted a national survey in 2004 to collect information on the types of medical equipment that is...
-
M303.4 Components (Section-by-Section Analysis)
The MDE NPRM proposed to require diagnostic equipment used by patients seated in a wheelchair which has components that are used to examine specific body parts, be capable of examining those...
-
5. Exception from the Requirements of M302 for Weight Scales with Integral Seats
The remaining commenter, a medical association, noted that it was unaware of any diagnostic equipment with a folding seat, but asserted that if patients can use the equipment in wheelchairs...
-
M303.3 Entry
Where there is a change in level at the entry to wheelchair spaces, the change in level shall comply with M303.3....
-
M303.3 Entry
Where there is a change in level at the entry to a wheelchair space, the change in level shall comply with M303.3....
-
M303.2.4.2.1 Depth
Wheelchair spaces shall include knee and toe clearance 17 inches (430 mm) deep minimum and 25 inches (635 mm) deep maximum....