Toe clearance shall be provided at a height of 9 inches (230 mm) above the floor to a depth of 6 inches (150 mm) maximum.
Search Results "Wheelchair Space (Medical Diagnostic Equipment)"
Commonly Searched Documents
-
M303.2.4.1 Toe Clearance
-
M303.2.4.2 Knee Clearance
Knee clearance shall be provided at a depth of 11 inches (280 mm) minimum and 25 inches (635 mm) maximum at 9 inches (230 mm) above the floor and at a depth of 8 inches (205 mm)...
-
M303.3.1 Vertical (Section-by-Section Analysis)
The MDE NPRM proposed that for equipment with a change in level at the entry to the wheelchair space, level changes of up to ¼ inch high are permitted to be vertical....
-
M303.2.6 Edge Protection (Section-by-Section Analysis)
(Significant Changes – Edge Protection), the final rule requires platforms with wheelchair spaces that are raised more than 1½ inches in height to provide a minimum 2-inch-high edge protection...
-
M303.3.2 Beveled (Section-by-Section Analysis)
The MDE NPRM proposed that for equipment with a change in level at the entry to the wheelchair space, level changes greater than ¼ inch but not greater than ½ inch would be required to be...
-
11B-805.4.2 Equipment
Clear space complying with Section 11B-305.2 shall be provided as required for specific equipment....
-
11B-805.4.2 Equipment
Clear space complying with Section 11B-305.2 shall be provided as required for specific equipment....
-
11B-805.4.2 Equipment
Clear space complying with Section 11B-305.2 shall be provided as required for specific equipment....
-
5.8.1.1 Clear Platform Size Recommendation
The Wheeled Mobility Anthropometry Project recommends this platform dimension.iv These study findings noted that a 36-inch length accommodates standard manual and power wheelchairs and most...
-
Wheelchair Spaces: ADA Standard Section 802.1
Wheelchair spaces shall comply with 802.1....
-
5.8.2.4 Unobstructed Floor Space Recommendation
Rationale for the recommendation The overall knee and toe clearance and unobstructed floor space dimensions dictate how close a patient in a wheelchair can get to the mammography equipment...
-
M303.2.2 Width
Wheelchair spaces shall be 36 inches (915 mm) wide minimum....
-
M303.2.4 Knee and Toe Clearance
Wheelchair spaces shall include knee and toe clearance complying with M303.2.4....
-
M304 Diagnostic Equipment Used by Patients in Standing Position
in a wheelchair)....
-
4. Organization of Technical Criteria
Medical diagnostic equipment is typically designed to support patients in certain positions....
-
5.9.2 Standing Supports on Ramped Entry Raised Platforms with Wheelchair Spaces
5.9.2 Standing Supports on Ramped Entry Raised Platforms with Wheelchair Spaces....
-
5.8.1.1 Clear Platform Size
5.8.1.1 Clear Platform Size. The platform should be a minimum clear width of 32-inches and a minimum clear length of 40-inches.
-
Base Support [No specific provision—affects M303.2.4]
It does obstruct the floor space in front of the gantry and, thus, may limit how close a wheelchair can get to the equipment....
-
M101.6 Dimensional Tolerances (Section-by-Section Analysis)
In the preamble of the MDE NPRM, the Access Board sought public input in question five on available information or resources concerning conventional industry tolerances for medical diagnostic...
-
M306.1 General (Section-by-Section Analysis)
The MDE NPRM proposed that, where diagnostic equipment communicates instructions or other information to the patient, the instructions or information must be provided in at least two of...
-
5.9.2 Standing Supports on Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations
5.9.2 Standing Supports on Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations...
-
1. Width and Depth of Wheelchair Spaces
The MDE NPRM proposed to require diagnostic equipment to have a wheelchair space that is at least 36 inches wide (proposed M303.2.2)....
-
M303.2.6 Edge Protection
Where wheelchair spaces are provided on a platform raised more than 1 ½ inches (38 mm) in height, edge protection 2 inches (51 mm) high minimum measured from the surface of the platform...
-
Other Sources
The Access Board used anthropometric data and other standards for the width of transfer surfaces on diagnostic equipment used by patients in a seated position (see M302.2.2), height of breast...