5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations The Committee recommends standing supports located on both sides of platforms,...
Search Results "Wheelchair Space (Medical Diagnostic Equipment)"
-
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations
-
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations The Committee recommends standing supports located on one side of the platform,...
-
M303.3.3.3 Clear Width
The clear width of ramp runs shall be 36 inches (915 mm) minimum.
-
M303.3.3.1 Running Slope
Ramp runs shall have a running slope not steeper than 1:12. EXCEPTION: A running slope not steeper than 1:8 shall be permitted for ramp runs with a maximum height of 2 ½ inches...
-
M303.3.3.4 Edge Protection
Ramps with drop offs ½ inch (13 mm) or greater shall provide edge protection 2 inches (50 mm) high minimum on each side with a drop off.
-
M303.3.3.5 Handrails
Ramps with a rise greater than 6 inches (150 mm) shall provide handrails on both sides.
-
M303.3.3.5 Handrails
Ramps with a rise greater than 6 inches (150 mm) shall provide handrails on each side.
-
M303.3.3.3 Clear Width
The clear width of ramp runs shall be 36 inches (915 mm) minimum.
-
M303.3.3.1 Running Slope
Ramp runs shall have a running slope not steeper than 1:12.
-
M201.1 Enforcing Authority
The proposed standards do not include scoping requirements that specify the minimum number of types of accessible diagnostic equipment required in different types of health care facilities...
-
M101.2 Application
Sections M301 through M304 shall be applied to diagnostic equipment, based on the patient positions that the equipment supports, during patient transfer and diagnostic use....
-
5. Edge Protection
The Access Board sought public input with question 30 in the MDE NPRM preamble, on whether there is diagnostic equipment with wheelchair spaces on raised platforms that does not provide...
-
Figure M5: Overview of M303 Diagnostic Equipment for Use by a Patient While Seated in a Wheelchair and 304 Diagnostic Equipment for Use by a Patient in a Standing Position (Mammography Machine Example)
Where the primary use of the equipment supports more than one patient position, it must comply with the requirements for each use. ...
-
5.8.2 Knee and Toe Clearance Under Breast Platform
The overall knee and toe space should be increased to a minimum of 28 inches deep, from the initially proposed 25 inch absolute dimension; increasing the knee clearance directly underneath...
-
5.8.2.3 Overall Knee and Toe Clearance
The overall knee and toe space should be increased to a minimum of 28 inches....
-
M303.2.5 Surfaces
Wheelchair space surfaces shall not slope more than 1:48 in any direction....
-
M303.2.5 Surfaces
Wheelchair space surfaces shall not slope more than 1:48 in any direction....
-
5.8.2.3 Overall Knee and Toe Clearance Recommendation
5.8.2.3 Overall Knee and Toe Clearance Recommendation The Committee recommends increasing the overall knee and toe space to 28 inches minimum....
-
Chapter M3 Technical Criteria
Chapter M3 provides technical criteria for accessible diagnostic equipment based on the patient positions that the equipment is designed to support, including equipment used by patients...
-
A. Chapter 2: M201 Scoping
In the final rule, Chapter 2 establishes that the enforcing authority will determine the number and types of diagnostic equipment to which the MDE Standards will apply....
-
5.8.2.7 Base Support Allowance
Base support is permitted to obstruct the clear floor space if it fits within the allowed base support volume. Base supports can be a maximum of 1½ inches high....
-
M101.6 Dimensional Tolerances
Dimensions are subject to conventional industry tolerances for manufacturing processes, material properties, and field conditions.
-
Chapter 2: Scoping
[See subsections ...]
-
M102.1 Defined Terms
For the purpose of the MDE Standards, the following terms have the indicated meaning: