M303.2.4 Equipment Clearances (Section-by-Section Analysis)

The MDE NPRM proposed knee and toe clearance for diagnostic equipment used by patients seated in wheelchairs to allow for components in the wheelchair space which the patient could approach successfully to use for its intended diagnostic purpose. The proposed requirements for equipment clearances paralleled the knee and toe clearance requirements from the 2004 ADA and ABA Accessibility Guidelines. The proposed rule provided one additional requirement for breast platforms on mammography equipment, proposing the knee and toe clearance under a breast platform to be 25 inches deep (proposed M303.2.4). The MDE NPRM preamble sought input with question 34 on whether the dimensions recommended by the Wheeled Mobility Anthropometry Project should be adopted.10 Three commenters responded. A manufacturer asserted that adopting a different requirement than what is already required under existing accessibility guidelines and standards would cause confusion and increase costs. A medical association asserted that to the best of their knowledge, imaging equipment already meets the Wheeled Mobility Anthropometry Project recommendations. The final commenter, a state agency concerned with accessibility, recommended adopting the new Wheeled Mobility Anthropometry Project recommendations. The MDE Advisory Committee only provided recommendations pertaining to the knee and toe clearance for mammography equipment.

The Access Board has determined that mammography equipment presents a unique challenge. Mammography equipment contains breast platforms which patients seated in wheelchairs must approach, and successfully maneuver their lower body under the platform enough to allow their chest to be flush with the leading edge of the platform. A separate set of equipment clearance requirements is necessary to address the unique positioning at mammography equipment. Therefore, in the final rule the Access Board has separated out the knee and toe clearance requirements into two provisions; breast platforms and other equipment. Breast platform requirements address the knee and toe clearance requirements for mammography equipment which is usable by patients seated in a wheelchair and is discussed in Section IV.C.2. (Significant Changes – Equipment Clearances for Breast Platforms). All other diagnostic equipment used by patients seated in a wheelchair must comply with the other equipment clearances requirements.

For all other equipment, the Access Board has decided to retain the original requirements in the proposed rule for knee and toe clearance. The Access Board is not persuaded to adopt the Wheeled Mobility Anthropometry Project recommendations for knee and toe clearances at this time. These recommendations represent a significant departure from the 2004 ADA and ABA Accessibility Guidelines. Therefore, the Board has elected in the final rule to retain the proposed provisions in the NPRM for knee and toe clearance for other equipment (M303.2.4.2). Due to the reorganization of the equipment clearances provision in the final rule, the knee and toe clearance requirements for the other equipment section have been renamed depth and height and relocated to M303.2.4.2. In addition, the Access Board has made an editorial change to the toe height requirement to clarify that the measurement is taken from the toe end of the wheelchair space.

 10 The Wheeled Mobility Anthropometry Project recommended a toe clearance that is 5 inches deep maximum at 14 inches above the floor and a knee clearance that is 12 inches deep minimum at 28 inches above the floor.

[MORE INFO...]

*You must sign in to view [MORE INFO...]