The length of gripping surfaces on horizontal standing supports shall be 4 inches (100 mm) minimum except for diagnostic equipment containing a wheelchair space which shall comply...
Search Results "Patient in Side-Lying Position"
-
M305.3.1.1 Horizontal Position
-
M305.3.2.1 Horizontal Position
The height of the top of the gripping surface on horizontal standing supports shall be 34 inches (865 mm) minimum and 38 inches (965 mm) maximum above the standing surface.
-
Visible Indicators: ADA Standard Section 407.4.8.1
Section 407.4.8.1 covers the current ADA Standards for visible elevator car position indicators.
-
6.3(5)
(5) Each bedroom shall have a door that complies with 4.13.
-
11B-604.5.1 Side wall
The side wall grab bar shall be 42 inches (1067 mm) long minimum, located 12 inches (305 mm) maximum from the rear wall and extending 54 inches (1372 mm) minimum from the rear wall with...
-
11B-604.5.1 Side wall
The side wall grab bar shall be 42 inches (1067 mm) long minimum, located 12 inches (305 mm) maximum from the rear wall and extending 54 inches (1372 mm) minimum from the rear wall with...
-
M302.1 General (Section-by-Section Analysis)
The MDE NPRM proposed that all diagnostic equipment used by patients in a seated position must comply with the technical requirements of proposed section M302....
-
11B-604.8.1.1.2 Maneuvering space with side-opening door
In a wheelchair accessible compartment with a door located in the side wall or partition, either in-swinging or out-swinging, a minimum 60 inches (1524 mm) wide and 60 inches (1524 mm) deep...
-
11B-604.8.1.1.2 Maneuvering space with side-opening door
In a wheelchair accessible compartment with a door located in the side wall or partition, either in-swinging or out-swinging, a minimum 60 inches (1524 mm) wide and 60 inches (1524 mm) deep...
-
6.4 PATIENT TOILET ROOMS
Provide each patient bedroom that is required to be accessible with an accessible toilet room that complies with 4.22 or 4.23....
-
M302.3.2 Leg Supports
Where stirrups are provided, leg supports shall also be provided and comply with M305.4.
-
M302.3 Supports
Transfer supports, leg supports and reclining surfaces shall comply with M302.3.
-
M302.2 Transfer Surface
A transfer surface shall be provided and shall comply with M302.2.
-
M302.2 Transfer Surface (Section-by-Section Analysis)
This is an introductory section.
-
M302.3 Supports (Section-by-Section Analysis)
This is an introductory section. An editorial change was made to this section as a result of the change in M302.3.2, described below, to replace the word “stirrups” with the term...
-
M302.2.2 Size
The transfer surface shall be 21 inches (610 mm) wide minimum and 15 inches (381 mm) deep minimum. Advisory M302.2.2 Size. The size requirements in this section apply only to the...
-
M302.2 Transfer Surface
A transfer surface shall be provided and shall comply with M302.2.
-
M302.3 Supports
Transfer supports, armrests, and reclining surfaces shall comply with M302.3.
- The Association of Safe Patient Handling Professionals
-
11B-223.1.1.3 Toilet and bathing facilities
Toilet/bathing rooms which are part of patient bedrooms added or altered and required to be accessible shall comply with Section 11B-805.2.4....
-
11B-223.1.1.3 Toilet and bathing facilities
Toilet/bathing rooms which are part of patient bedrooms added or altered and required to be accessible shall comply with Section 11B-805.2.4....
-
Accessible Patient Rooms
There are certain types of patient rooms, such as those provided in critical or intensive care units where patients who are critically ill are immobile or confined to beds and thus generally...
-
Facilities Not Specializing in Treating Conditions That Affect Mobility: ADA Standard Section 223.2.1
Section 223.2.1 covers scoping requirements in the current ADA Standards for facilities not specializing in treating conditions that affect mobility.
-
M302.3.3 Head and Back Support
Where the diagnostic equipment is used in a reclined position, head and back support shall be provided and shall comply with M305.5....