Figure M1: Overview of M301 Diagnostic Equipment for Use by a Patient in a Supine, Prone, or Side-Lying Position
-
5.1 Transfer Surface Height
5.1 Transfer Surface Height
-
5.1.1 Transfer Surface Height Adjustability
5.1.1 Transfer Surface Height Adjustability. The transfer surface should have an adjustable height in small virtually continuous increments.
-
5.1.2 Transfer Surface High Height
5.1.2 Transfer Surface High Height. The transfer surface should have an adjustable height range with an upper height measuring at least 25 inches.
-
5.1.3 Transfer Surface Low Height
5.1.3 Transfer Surface Low Height. No agreement on final low height
-
5.1.4 Transfer Surface Height Measurement
5.1.4 Transfer Surface Height Measurement. The transfer surface height should be measured from the floor to the highest point on the seat in an uncompressed state, inclusive of...
-
5.2 Transfer Surface Size for Equipment Used by Patients in Supine, Prone, or Side-Lying Position
5.2 Transfer Surface Size for Equipment Used by Patients in Supine, Prone, or Side-Lying Position....
-
5.2.1 Transfer Surface Size
5.2.1 Transfer Surface Size
-
5.2.1.1 Transfer Surface Width
The transfer surface should be 28 inches wide minimum for equipment used by patients in a supine, side-lying, or prone position....
-
5.2.1.2 Transfer Surface Depth
The transfer surface should be 17 inches deep minimum for all equipment used by patients in a supine, side-lying, or prone position with the exception of imaging equipment....
-
5.2.1.3 Transfer Surface Size for Stretchers
The transfer surface should be 28 inches wide minimum and 17 inches deep minimum located on both long sides of stretchers....
-
5.2.1.4 Transfer Surface Size for Imaging Equipment
The transfer surface should be located such that the long dimension of 28 inches is located parallel to the patient scanning/imaging side of the table....
-
5.3 Transfer Sides
5.3 Transfer Sides....
-
5.3.1 Permitted Obstructions to Transfer Sides for Equipment Used by Patients in a Supine, Prone, Side-Lying or Seated Position
5.3.1 Permitted Obstructions to Transfer Sides for Equipment Used by Patients in a Supine, Prone, Side-Lying or Seated Position....
-
5.3.2 Transfer Sides for Equipment Used by Patients in a Supine, Prone, or Side-Lying Position
5.3.2 Transfer Sides for Equipment Used by Patients in a Supine, Prone, or Side-Lying Position....
-
5.3.2.1 Transfer Sides for Stretchers
5.3.2.1 Transfer Sides for Stretchers. The transfer surface shall be located to provide the ability to transfer from a mobility device onto both long sides of the surface....
-
5.3.2.2 Transfer Sides for Imaging Equipment
5.3.2.2 Transfer Sides for Imaging Equipment. The transfer surface should be provided on at least one long side of the table....
-
5.4 Transfer Supports
5.4 Transfer Supports.
-
5.4.1 Transfer Support Location for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.1 Transfer Support Location for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
5.4.1.1 Transfer Support Location for Stretchers
The transfer surface for stretchers should be oriented along the long side of the surface....
-
5.4.1.2 Transfer and Positioning Support Location for Imaging Equipment
5.4.1.2 Transfer and Positioning Support Location for Imaging Equipment....
-
5.4.2 Transfer Support Length for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.2 Transfer Support Length for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
5.4.2.1 Transfer Support Length for Stretchers
5.4.2.1 Transfer Support Length for Stretchers. Transfer supports on stretchers should be a minimum of 15 inches long.
-
5.4.2.2 Transfer and Positioning Support Length for Imaging Equipment with Transfer Surfaces
5.4.2.2 Transfer and Positioning Support Length for Imaging Equipment with Transfer Surfaces....