Patient Load Rating: 450 lb. (204 kg) Base Height: Minimum 19" (48.3 cm) Maximum 40" (101.6 cm) Foot Section Height: With foot section retracted 15.5" to 54"...
Search Results "MDE: Medical Diagnostic Equipment"
Commonly Searched Documents
- Midmark 647 Barrier-Free® Power Podiatry Procedures Chair
- Midmark 630 HUMANFORM™ Procedures Table
-
5.8.3 Breast Platform Recommendations
clearance at the breast platform (see 5.8.2 Knee and Toe Clearance Recommendations under Breast Platforms), the height of the breast platform is also critical to ensuring that mammography equipment...
-
M305.3.1 Length
Horizontal standing support gripping surfaces shall comply with M305.3.1.1, horizontal standing support gripping surfaces on diagnostic equipment containing a wheelchair space shall comply...
-
5.8.2.4 Unobstructed Floor Space Recommendation
Rationale for the recommendation The overall knee and toe clearance and unobstructed floor space dimensions dictate how close a patient in a wheelchair can get to the mammography equipment...
-
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces. Standing supports should be located on both sides of platforms, a minimum of 34 inches...
-
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations The Committee recommends standing supports located on one side of the...
-
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations The Committee recommends standing supports located on both sides of...
-
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces. Standing supports should be located on one side of the platform, integrated into the platform...
-
5.8.2.7 Base Support Allowance Recommendation
It does obstruct the floor space in front of the gantry and, thus, may limit how close a wheelchair can get to the equipment....
-
M303.3.3.3 Clear Width
The clear width of ramp runs shall be 36 inches (915 mm) minimum.
-
M303.3.3.1 Running Slope
Ramp runs shall have a running slope not steeper than 1:12. EXCEPTION: A running slope not steeper than 1:8 shall be permitted for ramp runs with a maximum height of 2 ½ inches...
-
M303.3.3.4 Edge Protection
Ramps with drop offs ½ inch (13 mm) or greater shall provide edge protection 2 inches (50 mm) high minimum on each side with a drop off.
-
M303.3.3.5 Handrails
Ramps with a rise greater than 6 inches (150 mm) shall provide handrails on both sides.
-
M303.3.3.5 Handrails
Ramps with a rise greater than 6 inches (150 mm) shall provide handrails on each side.
-
M303.3.3.3 Clear Width
The clear width of ramp runs shall be 36 inches (915 mm) minimum.
-
M303.3.3.1 Running Slope
Ramp runs shall have a running slope not steeper than 1:12.
-
5.8.3.1 Breast Platform Height Recommendations
According to industry, equipment currently manufactured ranges anywhere between 25 and 28 inches for the lowest measurement of the breast platform....
-
Figure M4: Overview of M303 Diagnostic Equipment for Use by a Patient While Seated in a Wheelchair and M304 Diagnostic Equipment for Use by a Patient in a Standing Position (Weight Scale Example)
Where the primary use of the equipment supports more than one patient position, it must comply with the requirements for each use. ...
-
7. Diagnostic Imaging Equipment: Accessibility Considerations
Diagnostic Imaging Equipment: Accessibility Considerations Section 4 introduces the complex considerations raised when considering accessibility standards, such as transfer surface heights...
-
5.8.1.2.1 Ramped Entry Slope Recommendations
The space constraints affect the desirability of accessible scales since space is often expensive and tight in many medical facilities....
-
M303.2.4.1 Breast Platforms
Wheelchair spaces beneath breast platforms shall comply with M303.2.4.1.
-
M303.2.4.1.1 Depth
Wheelchair spaces shall include knee and toe clearance 25 inches (635 mm) deep minimum and 28 inches (710 mm) deep maximum.
-
M303.2.4.1.2 Height
Wheelchair spaces shall include toe clearance 9 inches (230 mm) high minimum above the floor measured to a depth of 6 inches (150 mm) maximum from the toe end of the wheelchair...