Preamble Discussion: The Access Board is considering whether 1½ inches minimum clearance around the gripping surface would be appropriate for transfer supports on diagnostic equipment used by patients...
Search Results "Patient in Side-Lying Position"
-
5.4.7 Transfer Support Gripping Surface Clearance Recommendation for M301 and M302
-
5.4.8 Transfer Support Gripping Surface Hazard Recommendation for M301 and M302
Rationale for the recommendation Gripping surface configurations must provide an effective and safe surface for patients to hold onto....
-
5.4.9 Interruptions along Transfer Support Gripping Recommendation for M301 and M302
5.4.9 Interruptions along Transfer Support Gripping Recommendation for M301 and M302 Description: Transfer supports may contain elements to provide structural support or prevent patient...
-
5.4.10 Transfer Support Fittings Recommendations for M301 and M302
The Committee recommends transfer supports not rotate when locked in place for patient transfer or use....
-
5.6.1 Stirrups Provision for M301
The proposed provision ensures a method of supporting, positioning, and securing the patient’s legs. NPRM Proposed Provision: M301.3.2 Stirrups....
-
5.7.1 Lift Compatibility Clearance in Base for M301
portable floor lifts to be effective, they must deploy at diagnostic equipment such that the boom of the lifts can maneuver far enough over the equipment to safely lower and raise the patient...
-
Transfer Surface Size – M301.2.2 Supine, Prone, or Side-Lying Position
Transfer Surface Size – M301.2.2 Supine, Prone, or Side-Lying Position The issues were discussed and decided in the full committee meeting as follows: depth of 17 inches minimum and width...
-
Height – M301.2.1 & M302.2.1
subcommittee discussed the increments for adjustability and decided that infinite adjustability is the appropriate standard as both electronic and manual pump chairs have ability to stop at any position...
-
17 Inch Low Height
A 17 inch height allows for safer, easier transfers of patients whose wheelchair height is at the low end of those ranges....
-
18 Inch Low Height
18 Inch Low Height As our discussions progressed on the low height, whether it should be 17 inches, 18 inches or 19 inches; a compromise position of 18 inches was suggested....
-
19 Inch Low Height
support surfaces are contoured or upholstered for patient comfort or to support patient positioning during diagnostic procedures, the height of the transfer surface measured from the floor...
-
How to Measure Transfer Surface Height
The corner radii allow for closer wheelchair positioning to facilitate independent transfer by minimizing gaps....
-
Transfer Supports Location and Configuration M301.3 and M302.3
The length of the transfer support shall be a minimum of 15 inches long and positioned so that the transfer support overlaps the length of the transfer surface by 80% at a maximum distance...
-
Gripping Surface Cross Section and Clearances M301, M302, and M305.2
Gripping Surface Cross Section and Clearances M301, M302, and M305.2 Gripping surface for transfer supports shall be free of sharp or abrasive elements and shall have rounded...
-
M301.2.1 Transfer Surface Height
Also note the equipment imaging components on the opposite side of where the patient transfers....
-
M301.2.2 Transfer Surface Size
The minimum size of the transfer surface is 28 inches wide minimum and 21 inches deep minimum and is located such that the long dimension of 28 inches is be located parallel to the patient...
-
M301.2.3 Transfer Surface Transfer Sides
The option to transfer will be required only on one long side of the patient scanning/imaging table and not at the “foot” or “head” end of such surfaces....
-
M301.3.1 Transfer Supports
B) For transfer depths greater than 24 inches a positioning support will be located opposite the transfer side....
-
6. Transfer Surface Height
Given that diagnostic imaging tables are in integral part of the exam and required for accurate results, their design to support very heavy patients, and in some types the existence of imaging...
-
M301.2.1, M302.2.1 Transfer Height
to the uncompressed height of the patient support surface....
-
M301.2.2, M302.2.2 Transfer Size
This transfer surface is located on the longitudinal (long) side of the stretcher—either right or left side Rationale: Based upon full committee discussion and harmonization with Exam...
-
M301.2.3 Transfer Sides
The location of the Transfer Side will be located on the patient right or patient left at roughly midway between the foot end and head end....
-
M301.3 Transfer Supports
Transfer Supports Position • Sub-Committee Recommendation: Located longitudinally within the long side of the of the transfer surface....
-
M301.3.2 Stirrups
M301.3.2 Stirrups Proposed requirement: Where stirrups are provided, they shall provide a method of supporting, positioning, and securing the patient’s legs....