This provision will coincide with the seat area design of most examination chairs. NPRM Proposed Provision: M302.2.3 Transfer Sides....
Search Results "Patient in Seated Position"
-
5.3.3 Transfer Sides Recommendations for M302
-
5.3.3.1 Transfer Sides Recommendation for Exam Chairs with Fixed Footrests
The current design allows only one long side for transfer, which limits some patient transfers where a patient can use only one side of the body due to paralysis on one side or other such...
-
5.4 Transfer Supports Recommendations
5.4 Transfer Supports Recommendations
-
5.4.1 Transfer Support Location for M301 and M302
Transfer supports shall be located within reach of the transfer surface and shall not obstruct transfer onto or off the surface when in position....
-
5.4.2 Transfer Support Length for M301 and M302
The purpose of a minimum length is to ensure that there is a minimum gripping surface available to support positioning and repositioning during a transfer....
-
5.4.3 Transfer Support Height for M301 and M302
The purpose of a minimum height is to ensure that the gripping surface is at an adequate height to facilitate positioning during a transfer....
-
5.4.4 Transfer Support Distance from Transfer Surface Recommendations for M301 and M302
in a supine, prone or side-lying position, and diagnostic equipment used by patients in a seated position: At least one transfer support would be provided on the side of the transfer...
-
5.4.5 Transfer Support Position Recommendation for M301 and M302
5.4.5 Transfer Support Position Recommendation for M301 and M302 Description: Transfer support position refers to the placement of the gripping surface within the minimum and maximum...
-
5.4.6 Transfer Support Gripping Surface Cross Section and Clearances for M301, M302, and M305.2
The Access Board is considering whether the above cross section dimensions would be appropriate for the gripping surfaces of transfer supports on diagnostic equipment used by patients in...
-
5.4.7 Transfer Support Gripping Surface Clearance Recommendation for M301 and M302
Preamble Discussion: The Access Board is considering whether 1½ inches minimum clearance around the gripping surface would be appropriate for transfer supports on diagnostic equipment used by patients...
-
5.4.8 Transfer Support Gripping Surface Hazard Recommendation for M301 and M302
Rationale for the recommendation Gripping surface configurations must provide an effective and safe surface for patients to hold onto....
-
5.4.9 Interruptions along Transfer Support Gripping Recommendation for M301 and M302
5.4.9 Interruptions along Transfer Support Gripping Recommendation for M301 and M302 Description: Transfer supports may contain elements to provide structural support or prevent patient...
-
5.4.10 Transfer Support Fittings Recommendations for M301 and M302
The Committee recommends transfer supports not rotate when locked in place for patient transfer or use....
-
5.5.1 Armrest Provision for M302
5.5.1 Armrest Provision for M302 Description: Diagnostic equipment used by patients in a seated position such as exam chairs typically have armrests....
-
5.7 Lift Compatibility Recommendations
5.7 Lift Compatibility Recommendations
-
Transfer Surface Size – M302.2.2 Seated Position
Transfer Surface Size – M302.2.2 Seated Position The issue of seat depth was discussed and decided during the full committee meetings as follows: depth of 17 inches minimum....
-
Height – M301.2.1 & M302.2.1
subcommittee discussed the increments for adjustability and decided that infinite adjustability is the appropriate standard as both electronic and manual pump chairs have ability to stop at any position...
-
17 Inch Low Height
The height was measured as the vertical distance from the floor to the lowest point of the seating surface of the mobility device, while the occupant was seated in the device....
-
18 Inch Low Height
18 Inch Low Height As our discussions progressed on the low height, whether it should be 17 inches, 18 inches or 19 inches; a compromise position of 18 inches was suggested....
-
19 Inch Low Height
, shower seats, dressing benches).”7 In addition, the Access Board recommended, “Where patient support surfaces are contoured or upholstered for patient comfort or to support patient positioning...
-
How to Measure Transfer Surface Height
Manufacturers designed the shape of the seat for these tables to meet both patient accessibility and clinical needs, resulting in a complex, contoured shape....
-
Transfer Supports Location and Configuration M301.3 and M302.3
The length of the transfer support shall be a minimum of 15 inches long and positioned so that the transfer support overlaps the length of the transfer surface by 80% at a maximum distance...
-
Gripping Surface Cross Section and Clearances M301, M302, and M305.2
Gripping Surface Cross Section and Clearances M301, M302, and M305.2 Gripping surface for transfer supports shall be free of sharp or abrasive elements and shall have rounded...
-
Armrests M302.3.2
Armrests M302.3.2 Armrests are recommended but not required. If provided armrests cannot interfere with transfer supports during the transfer. Dual transfer support/armrest...