Transfer supports shall comply with M305.2.
Search Results "Patient in Seated Position"
-
M305.2 Transfer Supports
-
M305.2 Transfer Supports
Transfer supports shall comply with M305.2.
-
M305.2.1 Location
Transfer supports shall comply with M305.2.1.1 or M305.2.1.2 and shall be located 1 ½ inches (38 mm) maximum measured horizontally from the plane defined by the nearest edge of the...
-
M305.2.1 Location
Transfer supports shall be located within reach of the transfer surface and shall not obstruct transfer onto or off of the surface when in position....
-
M305.2.1.1 End Transfer Supports
Transfer supports for transfer surfaces complying with M301.2.3.1 and M302.2 shall be located on the short side (length) opposite the transfer side.
-
M305.2.1.2 Side Transfer Supports
Transfer supports for transfer surfaces complying with M301.2.3.2 shall be capable of supporting transfer on each side of the transfer surface.
-
M305.2.2 Length
The length of transfer supports shall comply with M305.2.2.1 or M305.2.2.2.
-
M305.2.2.1 End Transfer Supports
Transfer supports shall be positioned along 13 ½ inches (345 mm) minimum of the depth of the transfer surface....
-
M305.2.2.2 Side Transfer Supports
Transfer supports for transfer surfaces complying with M301.2.3.2 shall be 28 inches (710 mm) long minimum and shall be positioned along the width of transfer surfaces....
-
M305.2.3 Fittings
Transfer supports shall not rotate within their fittings.
-
M305.2.3 Height
During use, the tops of transfer support gripping surfaces shall be 6 inches (150 mm) minimum and 19 inches (485 mm) maximum higher than the top of the associated uncompressed...
-
M305.2.4 Cross Section
Transfer supports shall have a cross section complying with 305.2.4.1 or 305.2.4.2.
-
M305.2.4.1 Circular Cross Section
Transfer supports with circular cross sections shall have an outside diameter of 1 ¼ inches (32 mm) minimum and 2 inches (51 mm) maximum.
-
M305.2.4.2 Non-Circular Cross Section
Transfer supports with non-circular cross sections shall have a cross-section dimension of 2 inches (51 mm) maximum and a perimeter dimension of 4 inches (100 mm) minimum and 4.8...
-
M305.2.5 Surface Hazards
Transfer supports and surfaces adjacent to transfer supports shall be free of sharp or abrasive components and shall have eased edges.
-
M305.2.6 Gripping Surface
Transfer support gripping surfaces shall be continuous along their length and shall not be obstructed along their tops or sides. The bottoms of transfer support gripping surfaces...
-
M305.2.7 Clearance
Clearance between the transfer support gripping surface and adjacent surfaces or obstructions shall be 1 ½ inches (38 mm) minimum.
-
M305.2.8 Fittings
Transfer supports shall not rotate within their fittings when in place for transfer.
-
M301 Diagnostic Equipment Used by Patients in Supine, Prone, or Side-Lying Position; M302 Diagnostic Equipment Used by Patients in Seated Position
patients in a seated position....
-
Transfer Surface (M301.2 and M302.2)
The transfer surface is the part of the diagnostic equipment onto which patients who use mobility devices or aids transfer when moving onto and off of the equipment (see defined terms in...
-
Transfer Surface Height (M301.2.1 and M302.2.1)
For many patients who use mobility devices, independent transfer is possible only if the height of the transfer surface is at or near the seat height of their mobility device....
-
Adjustable Height Range Considered
Many types of diagnostic equipment used by patients in a supine, prone, or side-lying position, and diagnostic equipment used by patients in a seated position currently provide adjustable...
-
Transfer Surface Size: Equipment Used by Patients in a Seated Position (M302.2.2)
Seats on diagnostic equipment used by patients in a seated position typically provide back and arm support for patient comfort and stability....
-
Transfer Sides (M301.2.3 and M302.2.3)
This is consistent with ANSI/AAMI HE 75 which recommends that “side rails, arm rests, leg supports . . . should be positioned, or able to be moved out of the way, so as not to interfere...