The force required to activate operable parts shall be 5 pounds (22.2 N) maximum.
Search Results "Equipment Console"
Commonly Searched Documents
-
M307.4 Operating Force
-
M307.4 Operating Force (Section-by-Section Analysis)
The MDE NPRM proposed to restrict the force required to activate operable parts to 5 pounds. The Access Board sought public input on this provision in question 42 on whether the...
-
M307.4 Operating Force
The force required to activate operable parts shall be 5 pounds (22.2 N) maximum.
-
3. Background
[See subsections ...]
-
5.8 Wheelchair Spaces for Diagnostic Equipment Used by Patients Seated in a Wheelchair
5.8 Wheelchair Spaces for Diagnostic Equipment Used by Patients Seated in a Wheelchair....
-
Compliance requirements in proposed standards
The proposed standards contain technical criteria for accessible medical diagnostic equipment....
-
5.4.4 Transfer Support Distance From Transfer Surface for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.4 Transfer Support Distance From Transfer Surface for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
5.4.3 Transfer Support Height for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.3 Transfer Support Height for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
- Senographe Essential eContrast
-
5.2.1.2 Transfer Surface Depth
The transfer surface should be 17 inches deep minimum for all equipment used by patients in a supine, side-lying, or prone position with the exception of imaging equipment....
-
Cassette Recordings (Advisory Guidance)
Cassette Recordings 9. Some persons who are blind or who have learning disabilities may require documentation on audio cassettes. Audio materials can be produced commercially or...
-
Large Print (Advisory Guidance)
Large Print 7. Persons with low vision may require documentation to be provided in large print. Large print documents can easily be produced using a scalable font from any good...
-
Electronic Text (Advisory Guidance)
Electronic Text 5. People who are blind or have low vision and who have access to computers may be able to use documents in electronic form. Electronic text must be provided in...
-
C201.5 Design, Development, and Fabrication
Telecommunications equipment manufacturers shall evaluate the accessibility, usability, and interoperability of ICT during its product design, development, and fabrication....
-
M301.4.2 Clearance Around Base
The base of the equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination surface...
-
M301.4.2 Clearance Around Base
The base of diagnostic equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination...
-
A. Chapter 1: Application and Administration (Section-by-Section Analysis)
The Access Board made a few editorial changes to some of the provisions, and added one provision M101.3 Existing Diagnostic Equipment, which is discussed below....
-
5.2.1.1 Transfer Surface Width Recommendations for M301
Currently, the only equipment existing with a 30-inch transfer surface width is bariatric equipment. At the Committee’s December 2012 meeting, Dr....
-
M301.2.1 Transfer Surface Height
Most tables for use with equipment for bores are height adjustable....
-
5.2.1 Transfer Surface Size Recommendations for M301
5.2.1 Transfer Surface Size Recommendations for M301 Description: The transfer surface is the part of the diagnostic equipment onto which patients who use mobility devices or aids transfer...
-
M101.2 Application
The standards shall be applied to diagnostic equipment based on the patient positions that the equipment is designed to support....
-
2. Equipment Clearances for Breast Platforms
Additionally, this commenter raised concerns that mammography equipment must have a stabilizing flange or foot at its base to prevent the equipment from tipping when the gantry is extended...
-
5.3.1 Permitted Obstructions to Transfer Sides for Equipment Used by Patients in a Supine, Prone, Side-Lying or Seated Position
5.3.1 Permitted Obstructions to Transfer Sides for Equipment Used by Patients in a Supine, Prone, Side-Lying or Seated Position....
-
Regulatory Flexibility Act
In particular, the certification noted that manufacturers of telecommunications equipment and customer premises equipment are required to comply with section 255 of the Telecommunications...