It does obstruct the floor space in front of the gantry and, thus, may limit how close a wheelchair can get to the equipment....
Search Results "Wheelchair Space (Medical Diagnostic Equipment)"
Commonly Searched Documents
-
5.8.2.7 Base Support Allowance Recommendation
-
3. Background
[See subsections ...]
-
M102.1 Defined Terms (Section-by-Section Analysis)
The MDE NPRM proposed definitions for enforcing authority, medical diagnostic equipment, operable parts, and transfer surface....
-
5.8.3 Breast Platform Recommendations
clearance at the breast platform (see 5.8.2 Knee and Toe Clearance Recommendations under Breast Platforms), the height of the breast platform is also critical to ensuring that mammography equipment...
-
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces....
-
M303.2.4.1 Breast Platforms
Wheelchair spaces beneath breast platforms shall comply with M303.2.4.1....
-
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces....
-
M303.2.4.1.1 Depth
Wheelchair spaces shall include knee and toe clearance 25 inches (635 mm) deep minimum and 28 inches (710 mm) deep maximum....
-
M303.2.4.1.2 Height
Wheelchair spaces shall include toe clearance 9 inches (230 mm) high minimum above the floor measured to a depth of 6 inches (150 mm) maximum from the toe end of the wheelchair space....
- Selenia Dimensions Mammography System
-
M101.2 Application
The standards shall be applied to diagnostic equipment based on the patient positions that the equipment is designed to support....
-
5.8.3.1 Breast Platform Height Recommendations
According to industry, equipment currently manufactured ranges anywhere between 25 and 28 inches for the lowest measurement of the breast platform....
-
5.8.2.5 Toe Height Recommendation
However, as the breast platform moves up and down, it may intrude on the toe space....
-
1.3.1 Overview of NPRM
The proposed standards contain minimum technical criteria to ensure that medical diagnostic equipment including examination tables, examination chairs, weight scales, mammography equipment...
-
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations The Committee recommends standing supports located on both sides of platforms,...
-
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces Recommendations The Committee recommends standing supports located on one side of the platform,...
-
5.8.2.8 Knee and Toe Clearance Recommendations Summary
toe clearance recommendations in sections 5.8.2.1 – 5.8.2.7 are summarized below, in Figure 5.8.2.8, which shows the measurements needed to determine if a certain piece of mammography equipment...
-
M101.3 Existing Diagnostic Equipment (Section-by-Section Analysis)
The MDE NPRM did not address when or how the MDE Standards would be applied to existing medical diagnostic equipment....
-
M303.3.3.3 Clear Width
The clear width of ramp runs shall be 36 inches (915 mm) minimum.
-
M303.3.3.1 Running Slope
Ramp runs shall have a running slope not steeper than 1:12. EXCEPTION: A running slope not steeper than 1:8 shall be permitted for ramp runs with a maximum height of 2 ½ inches...
-
M303.3.3.4 Edge Protection
Ramps with drop offs ½ inch (13 mm) or greater shall provide edge protection 2 inches (50 mm) high minimum on each side with a drop off.
-
M303.3.3.5 Handrails
Ramps with a rise greater than 6 inches (150 mm) shall provide handrails on both sides.
-
M303.3.3.5 Handrails
Ramps with a rise greater than 6 inches (150 mm) shall provide handrails on each side.
-
M303.3.3.3 Clear Width
The clear width of ramp runs shall be 36 inches (915 mm) minimum.