Tips for Weighing People Using an Accessible Scale
-
5.8 Wheelchair Spaces for Diagnostic Equipment Used by Patients Seated in a Wheelchair
5.8 Wheelchair Spaces for Diagnostic Equipment Used by Patients Seated in a Wheelchair....
-
5.8.1 Wheelchair Spaces on Raised Platforms
5.8.1 Wheelchair Spaces on Raised Platforms....
-
5.8.1.1 Clear Platform Size
5.8.1.1 Clear Platform Size. The platform should be a minimum clear width of 32-inches and a minimum clear length of 40-inches.
-
5.8.1.2 Entry to Wheelchair Spaces on Raised Platforms
5.8.1.2 Entry to Wheelchair Spaces on Raised Platforms....
-
5.8.1.2.1 Ramped Entry Slope
5.8.1.2.1 Ramped Entry Slope. Raised platforms should have a ramp with slopes that do not exceed the following: Rise 0 to 1 ½ inches – 1:2 Slope Rise >1 ½ to 2 ½ inches – 1:8...
-
5.8.1.2.2 Single Ramped Entry – Edge Protection on the Platform and Platform Sides
5.8.1.2.2 Single Ramped Entry – Edge Protection on the Platform and Platform Sides. A 2 inch high edge protection must be provided on the back of the platform, opposite the entry...
-
5.8.1.2.3 Double Ramped Entry – Edge Protection on the Platform Sides
5.8.1.2.3 Double Ramped Entry – Edge Protection on the Platform Sides. A minimum 2 inch high edge protection must be provided on both sides of the platform for double ramped entry...
-
5.8.1.2.4 Edge Protection on Platforms 1½ inches or less in Height
5.8.1.2.4 Edge Protection on Platforms 1½ inches or less in Height. Edge protection is not required on platforms 1 ½ inches or less in height.
-
5.8.2 Knee and Toe Clearance Under Breast Platform
The overall knee and toe space should be increased to a minimum of 28 inches deep, from the initially proposed 25 inch absolute dimension; increasing the knee clearance directly underneath...
-
5.8.2.1 Height of Breast Platform At Time of Measurement
5.8.2.1 Height of Breast Platform At Time of Measurement. All knee and toe clearances should be measured when the top of the breast platform is set at 34 inches above the floor.
-
5.8.2.2 Knee Clearance Depth at 27” Above the Ground
5.8.2.2 Knee Clearance Depth at 27” Above the Ground. The knee clearance should be increased at 27 inches above the ground to 18 inches minimum.
-
5.8.2.3 Overall Knee and Toe Clearance
The overall knee and toe space should be increased to a minimum of 28 inches....
-
5.8.2.4 Unobstructed Floor Space
5.8.2.4 Unobstructed Floor Space. The unobstructed floor space in front of the base support should be a minimum of 17 inches deep....
-
5.8.2.5 Toe Height
5.8.2.5 Toe Height. The toe height should be 18 inches minimum, measured above the ground when the top of the breast platform is 34 inches above the ground.
-
5.8.2.6 Clearance Depth at Toe Height Above the Ground
5.8.2.6 Clearance Depth at Toe Height Above the Ground. The toe clearance depth should be 22 inches minimum at toe height.
-
5.8.2.7 Base Support Allowance
Base support is permitted to obstruct the clear floor space if it fits within the allowed base support volume. Base supports can be a maximum of 1½ inches high....
-
5.8.3 Breast Platform
5.8.3 Breast Platform
-
5.8.3.1 Breast Platform Height
5.8.3.1 Breast Platform Height. The height of the top of the breast platform should be decreased to 26 inches above the floor. The upper height range for the breast platform was not...
-
5.9.2 Standing Supports on Ramped Entry Raised Platforms with Wheelchair Spaces
5.9.2 Standing Supports on Ramped Entry Raised Platforms with Wheelchair Spaces....
-
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces
5.9.2.1 Standing Supports on Single Ramped Entry Raised Platforms with Wheelchair Spaces....
-
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces
5.9.2.2 Standing Supports on Dual Ramped Entry Raised Platforms with Wheelchair Spaces....
-
Wheelchair Space
Space for a single wheelchair and its occupant....
-
5. Exception from the Requirements of M302 for Weight Scales with Integral Seats
The remaining commenter, a medical association, noted that it was unaware of any diagnostic equipment with a folding seat, but asserted that if patients can use the equipment in wheelchairs...