The MDE NPRM proposed to require transfer supports to be provided for use with transfer sides on diagnostic equipment used by patients in the supine, prone, or side-lying position, and that...
Search Results "Positioning Support"
Commonly Searched Documents
-
M301.3.1 Transfer Supports (Section-by-Section Analysis)
-
Figure M1: Overview of M301 Diagnostic Equipment for Use by a Patient in a Supine, Prone, or Side-Lying Position
[See subsections ...]
-
M305.2.2.1 End Transfer Supports
Transfer supports for transfer surfaces complying with M301.2.3.1 and M305.2.2.1 shall be 15 inches (380 mm) long minimum....
-
M301.3.2 Leg Supports
Where stirrups are provided, leg supports shall also be provided and shall comply with M305.4....
-
5.4.8 Transfer Support Gripping Surface Hazard for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.8 Transfer Support Gripping Surface Hazard for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
M305.2.2.2 Side Transfer Supports
Transfer supports for transfer surfaces complying with M301.2.3.2 shall be 28 inches (710 mm) long minimum and shall be positioned along the width of transfer surfaces....
-
M305.3.1 Length (Section-by-Section Analysis)
(Significant Changes – Standing Supports), the final rule requires that horizontal standing supports be positioned horizontally in relation to standing surfaces and retains the proposed...
-
M305.3.1 Length
The length of gripping surfaces for standing supports shall be based on the position of the standing supports in relation to the standing surfaces they serve....
-
M301.3.1 Transfer Supports
Transfer surfaces required by M301.2 shall provide transfer supports and shall comply with M305.2....
-
M301.3.1 Transfer Supports
Transfer supports shall be provided for use with the transfer sides required by M301.2.3 and shall comply with M305.2....
-
5.4.3 Transfer Support Height for M301 and M302
5.4.3 Transfer Support Height for M301 and M302 Description: Transfer support height applies to the vertical distance that the gripping area of the support is above the transfer surface...
-
Shooting Facilities with Firing Positions: ADA Standard Section 243
Section 243 covers scoping requirements in the current ADA Standards for shooting facilities and firing positions.
-
Shooting Facilities with Firing Positions: ADA Standard Section 1010
[See subsections ...]
-
5.4.7 Transfer Support Gripping Surface Clearance Recommendation for M301 and M302
to grasp the transfer support....
-
5.4.4 Transfer Support Distance from Transfer Surface Recommendations for M301 and M302
surface and can be effectively used to position during a transfer....
-
M305.1 General
Supports shall comply with M305....
-
M305.1 General
Supports shall comply with M305, as applicable....
-
M305.3.2 Height
The height of gripping surfaces for standing supports shall be based on the position of the standing supports in relation to the standing surfaces they serve....
-
Position: ADA Standard Section 703.3.2
Braille shall be positioned below the corresponding text. If text is multi-lined, braille shall be placed below the entire text....
-
11B-1010 Shooting facilities with firing positions
[See subsections ...]
-
11B-243 Shooting facilities with firing positions
[See subsections ...]
-
Back Support
It was also recommended that the criteria for back support, which were included in the specifications for bench size, be relocated into a separate provision specific to back support....
-
11B-703.3.2 Position
Braille shall be positioned below the corresponding text in a horizontal format, flush left or centered. If text is multi-lined, Braille shall be placed below the entire text....
-
602.3 Electronic Support Documentation