In the MDE NPRM the Access Board proposed to place the requirements for head and back support for diagnostic equipment used by patients in the supine, prone, or side-lying position...
Search Results "Adjusted Construction Cost"
Commonly Searched Documents
-
M301.3.3 Head and Back Support (Section-by-Section Analysis)
-
5.3.2 Transfer Sides Recommendations for M301
5.3.2 Transfer Sides Recommendations for M301 Description: The transfer side provision requires the transfer surface to be located to provide options to transfer onto the short...
-
5.3.2 Transfer Sides for Equipment Used by Patients in a Supine, Prone, or Side-Lying Position
5.3.2 Transfer Sides for Equipment Used by Patients in a Supine, Prone, or Side-Lying Position.
-
Gripping Surface Cross Section and Clearances M301, M302, and M305.2
Gripping Surface Cross Section and Clearances M301, M302, and M305.2 Gripping surface for transfer supports shall be free of sharp or abrasive elements and shall have rounded...
-
M301.3.1 Transfer Supports
Transfer supports shall be provided for use with the transfer sides required by M301.2.3 and shall comply with M305.2.
-
M301.3 Supports
Transfer supports, stirrups, and reclining surfaces shall comply with M301.3.
-
M301.3.3 Head and Back Support
Where the diagnostic equipment is used in a reclined position, head and back support shall be provided. Where the incline of the back support can be modified while in use, head and...
- Sears Brands, LLC Craftsman Digital Torpedo Level
-
Sec.36.401(d)(1)
(1) For purposes of this paragraph (d) --
-
§ 36.401(d)(1)
(1) For purposes of this paragraph (d) –
-
Utility Sinks
A utility sink is deeper than a standard restroom basin or kitchen sink and can be used to clean large pots or equipment. Usually, utility sinks are provided only in highly developed...
-
Are the elements of path of travel and accessible routes substantively the same (such as slope, vertical clearance, width of path of travel, etc.)?
Compliance is required to the maximum extent feasible without exceeding 20 percent of the cost of the work directly associated with the installation of EVCS (see Section 11B-202.4 Exception...
-
11B-201.1 Scope, 11B-201.2 Application based on building or facility use, 11B-202.2 Additions, 11B-202.3 Alterations
Each addition to an existing building or facility shall comply with the requirements for new construction and shall comply with Section 11B-202.4. 11B-202.3 Alterations....
-
Overview
March 15, 2012, is also the compliance date for using the 2010 Standards for new construction, alterations, program accessibility, and barrier removal....
- Magnuson Group RETTO Outdoor Waste Receptacle
- Landscape Forms Inc. Select Modular Steel Recycling Units
-
5.3.1 Permitted Obstructions to Transfer Sides for M301 and M302
Rationale for the recommendation Transfer supports provide handholds on adjustable medical equipment to facilitate transfers onto and off the equipment....
-
COSTS TO LOWER MINIMUM TABLE HEIGHT
COSTS TO LOWER MINIMUM TABLE HEIGHT To be useful for its intended purpose, examination tables and chairs must maintain a high height of at least 32 inches, so that the health care provider...
-
Exam Tables and Chairs
Individuals with mobility disabilities often need to use an adjustable-height table which, when positioned at a low height, allows them to transfer from a wheelchair....
- CalCasp Training: Access Plan Review
- Landscape Forms Inc. Presidio Trash Receptacle
-
M301.2.3.1 End Transfer Surface
End transfer surfaces shall be 28 inches (710 mm) wide minimum and 17 inches (430 mm) long minimum. EXCEPTION: Transfer surfaces for imaging equipment with bores shall be...
-
M301.2.3.2 Side Transfer Surface
Side transfer surfaces shall be 28 inches (710 mm) wide minimum and 28 inches (710 mm) long minimum. EXCEPTION: Transfer surfaces for imaging equipment with bores shall be...
-
M301.2.4 Unobstructed Transfer
Each transfer surface shall provide two unobstructed sides for patient transfer. EXCEPTIONS: 1. Obstructions no more than 3 inches (75 mm) deep shall be permitted to extend beyond...