Figure M3: Overview of M302 Diagnostic Equipment for Use by a Patient in a Seated Position
-
M301.2 Transfer Surface
A transfer surface shall be provided and shall comply with M301.2....
-
M301.2 Transfer Surface
A transfer surface shall be provided and shall comply with M301.2....
-
M301.2.1 Adjustability
Transfer surfaces shall be adjustable in height measured from the floor to the top of the uncompressed transfer surface and shall provide the following: A....
-
M301.2.1 Height
The height of the transfer surface during patient transfer shall be 17 inches (430 mm) minimum and 19 inches (485 mm) maximum measured from the floor to the top of the transfer surface....
-
M301.2.2 Sunset
The low transfer position height, Item A of M301.2.1, shall cease to have effect on [Insert date 5 years after date of publication in the Federal Register]....
-
M301.2.2 Size
The transfer surface shall be 30 inches (760 mm) wide minimum and 15 inches (381mm) deep minimum. Advisory M301.2.2 Size....
-
M301.2.3 Transfer Sides
The transfer surface shall be located to provide options to transfer from a mobility device onto one short side (depth) and one long side (width) of the surface....
-
M301.2.3 Size
The size of the transfer surface shall comply with M301.2.3.1 or M301.2.3.2. The size of transfer surfaces shall be measured from center points of their opposing sides....
-
M301.2.3.1 End Transfer Surface
End transfer surfaces shall be 28 inches (710 mm) wide minimum and 17 inches (430 mm) long minimum. ...
-
M301.2.3.2 Side Transfer Surface
Side transfer surfaces shall be 28 inches (710 mm) wide minimum and 28 inches (710 mm) long minimum....
-
M301.2.4 Unobstructed Transfer
Each transfer surface shall provide two unobstructed sides for patient transfer. EXCEPTIONS: 1....
-
M301.3.1 Transfer Supports
Transfer supports shall be provided for use with the transfer sides required by M301.2.3 and shall comply with M305.2....
-
M301.3.1 Transfer Supports
Transfer surfaces required by M301.2 shall provide transfer supports and shall comply with M305.2....
-
3.2 Getting on the Table
Additional helpful features include an extra-wide top surface and higher weight capacity for larger people, adjustable side rails, positioning straps, and foot and leg supports that can...
-
M302.2 Transfer Surface
A transfer surface shall be provided and shall comply with M302.2....
-
Seat width and depth
Seat width and depth: ADA-ABA currently requires a minimum entry point of 24 inches for transfer steps in play (Section 1008.3.1.1) [7] and pool areas (Section 1009.4.5) [7]. ...
-
M302.2 Transfer Surface
A transfer surface shall be provided and shall comply with M302.2....
-
M302.2.1 Height
The height of the transfer surface during patient transfer shall be 17 inches (430 mm) minimum and 19 inches (485 mm) maximum measured from the floor to the top of the transfer surface....
-
M302.2.1 Adjustability
Transfer surfaces shall be adjustable in height measured from the floor to the top of the uncompressed transfer surface and shall provide the following: A....
-
M302.2.2 Size
The transfer surface shall be 21 inches (610 mm) wide minimum and 15 inches (381 mm) deep minimum. Advisory M302.2.2 Size....
-
M302.2.2 Sunset
The low transfer position height, Item A of M302.2.1, shall cease to have effect on [Insert date 5 years after date of publication in the Federal Register]. [sic]...
-
M302.2.3 Size
Transfer surfaces shall be 21 inches (610 mm) wide minimum and 17 inches (430 mm) deep minimum....
-
M302.2.3 Transfer Sides
The transfer surface shall be located to provide options to transfer from a mobility device onto one short side (depth) and one long side (width) of the surface....