5.3.3.1 Transfer Sides for Exam Chairs with Fixed Footrests....
Search Results "Transfer Surface"
-
5.3.3.1 Transfer Sides for Exam Chairs with Fixed Footrests
-
5.4 Transfer Supports
5.4 Transfer Supports....
-
5.4.1 Transfer Support Location for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.1 Transfer Support Location for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
5.4.2 Transfer Support Length for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.2 Transfer Support Length for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
5.4.2.1 Transfer Support Length for Stretchers
5.4.2.1 Transfer Support Length for Stretchers. Transfer supports on stretchers should be a minimum of 15 inches long....
-
5.4.2.2 Transfer and Positioning Support Length for Imaging Equipment with Transfer Surfaces
5.4.2.2 Transfer and Positioning Support Length for Imaging Equipment with Transfer Surfaces....
-
5.4.3 Transfer Support Height for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.3 Transfer Support Height for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
5.4.4 Transfer Support Distance From Transfer Surface for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.4 Transfer Support Distance From Transfer Surface for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
5.4.4.1 Transfer Support Distance from Transfer Surface for Stretchers
5.4.4.1 Transfer Support Distance from Transfer Surface for Stretchers....
-
5.4.4.2 Transfer Support Distance from Transfer Surface for Imaging Equipment
5.4.4.2 Transfer Support Distance from Transfer Surface for Imaging Equipment. Supports (transfer or positioning) should be located 1 ½ inches maximum from the transfer surface....
-
5.4.5 Transfer Support Position for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.5 Transfer Support Position for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
5.4.5.1 Transfer and Positioning Support Position for Imaging Equipment
5.4.5.1 Transfer and Positioning Support Position for Imaging Equipment....
-
Transfer Surface
Part of diagnostic equipment onto which patients who use mobility devices or aids transfer when moving onto and off of the equipment....
-
Transfer Surface
Part of diagnostic equipment onto which patients who use mobility devices or aids transfer when moving onto and off of the equipment....
-
B. M301 Diagnostic Equipment Used by Patients in a Supine, Prone, or Side-lying Position and M302 Diagnostic Equipment Used by Patients in a Seated Position
Sections M301 and M302, which ensure that patients can transfer from their mobility devices onto the diagnostic equipment, share many technical requirements....
-
1. Transfer Surface
-
a. Transfer Surface Adjustability
The MDE NPRM proposed that the same transfer surface height range of 17 inches minimum to 19 inches maximum be applied to both diagnostic equipment used in the supine, prone, or side-lying...
-
(2) Adjustability: Minimum Low Transfer Height
These commenters explained the delay in diagnosis and treatment when patients are unable to transfer from their wheeled mobility device to the examination surface and are inadequately examined...
-
(3) Adjustability: Transfer Surface Intermediate Heights
In the MDE NPRM there was no requirement for the transfer surface to have intermediate transfer heights....
-
(4) Adjustability: Method of Measurement
used in the seated position, be taken from the floor to the top of the transfer surface (proposed M301.2.1 and M302.2.1, respectively)....
-
b. Transfer Surface Location
The MDE NPRM proposed the same location and transfer sides of the transfer surface for diagnostic equipment used by patients in the supine, prone, or side-lying position (M301) and diagnostic...
-
(1) Transfer Surface Location for Diagnostic Equipment Used in the Supine, Prone, or Side-lying Position
These orientations are now identified as an end transfer surface and side transfer surface....
-
(2) Transfer Surface Location for Diagnostic Equipment Used in the Seated Position
Transfer onto these types of diagnostic equipment must be made from either long side, similar to the side transfer surface described above....
-
c. Transfer Surface Size for Diagnostic Equipment Used by Patients in the Supine, Prone, or Side-lying Position
Those commenters supporting a smaller transfer surface raised concerns about the ability to transfer with a large surface preventing the patient from reaching transfer supports on the opposite...