Changes in level of ¼ inch (6.4 mm) high maximum shall be permitted to be vertical.
Search Results "Patient Seated in Wheelchair"
-
M303.3.1 Vertical
-
M303.3.1 Vertical
Changes in level of ¼ inch (6.4 mm) high maximum shall be permitted to be vertical.
-
M303.3.2 Beveled
Changes in level between ¼ inch (6.4 mm) high and ½ inch (13 mm) high maximum shall be beveled with a slope not steeper than 1:2.
-
M303.3.2 Beveled
Changes in level between ¼ inch (6.4 mm) high and ½ inch (13 mm) high maximum shall be beveled with a slope not steeper than 1:2.
-
M303.3.3 Ramped
Changes in level greater than ½ inch (13 mm) high shall be ramped and shall comply with M303.3.3.
-
M303.3.3 Ramped
Changes in level greater than ½ inch (13 mm) high shall be ramped and shall comply with M303.3.3.
-
M303.3.3.1 Running Slope
Ramp runs shall have a running slope not steeper than 1:12.
-
M303.3.3.1 Running Slope
Ramp runs shall have a running slope not steeper than 1:12. EXCEPTION: A running slope not steeper than 1:8 shall be permitted for ramp runs with a maximum height of 2 ½ inches...
-
M303.3.3.2 Cross Slope
The cross slope of ramp runs shall not be steeper than 1:48.
-
M303.3.3.2 Cross Slope
The cross slope of ramp runs shall not be steeper than 1:48.
-
M303.3.3.3 Clear Width
The clear width of ramp runs shall be 36 inches (915 mm) minimum.
-
M303.3.3.3 Clear Width
The clear width of ramp runs shall be 36 inches (915 mm) minimum.
-
M303.3.3.4 Edge Protection
Ramps with drop offs ½ inch (13 mm) or greater shall provide edge protection 2 inches (50 mm) high minimum on each side with a drop off.
-
M303.3.3.4 Edge Protection
Ramps with drop offs ½ inch (13 mm) or greater shall provide edge protection 2 inches (50 mm) high minimum on each side.
-
M303.3.3.5 Handrails
Ramps with a rise greater than 6 inches (150 mm) shall provide handrails on each side.
-
M303.3.3.5 Handrails
Ramps with a rise greater than 6 inches (150 mm) shall provide handrails on both sides.
-
M303.4 Components
Where components of diagnostic equipment are used to examine specific body parts, the components shall be capable of examining the body parts of a patient seated in a wheelchair....
-
M303.4 Components
Where components of diagnostic equipment are used to examine specific body parts, the components shall be capable of examining the body parts of a patient seated in a wheelchair....
-
M303.4.1 Breast Platforms
The height of the breast platform shall be 30 inches (760 mm) high minimum and 42 inches (1065 mm) high maximum above the floor when in use by a patient seated in a wheelchair....
-
M303.4.1 Breast Platform Adjustability
Breast platforms shall be continuously adjustable from a low height of 26 inches (660 mm) to a high height of 42 inches (1065 mm) above the floor.
-
Figure M4: Overview of M303 Diagnostic Equipment for Use by a Patient While Seated in a Wheelchair and M304 Diagnostic Equipment for Use by a Patient in a Standing Position (Weight Scale Example)
Where the primary use of the equipment supports more than one patient position, it must comply with the requirements for each use. ...
-
Figure M5: Overview of M303 Diagnostic Equipment for Use by a Patient While Seated in a Wheelchair and 304 Diagnostic Equipment for Use by a Patient in a Standing Position (Mammography Machine Example)
Where the primary use of the equipment supports more than one patient position, it must comply with the requirements for each use. ...
-
Folding Seats on Equipment Used by Patients Seated in a Wheelchair (M302.4 Exception)
M302.4 includes an exception for diagnostic equipment that is designed for use by patients seated in a wheelchair and provides a folding seat....
-
M303 Diagnostic Equipment Used by Patients Seated in Wheelchair
M303 provides technical criteria for diagnostic equipment used by patients seated in a wheelchair....