The MDE NPRM proposed that where there is a change in level at the entry of a wheelchair space that is greater than 1½ inches, the entry shall be ramped and have a running slope not steeper...
Search Results "Patient Seated in Wheelchair"
-
3. Exception to Ramp Running Slope
-
4. Breast Platform Adjustability
The MDE NPRM proposed to require diagnostic equipment used by patients seated in a wheelchair that have components which are used to examine specific body parts to be capable of examining...
-
5. Edge Protection
The Access Board sought public input with question 30 in the MDE NPRM preamble, on whether there is diagnostic equipment with wheelchair spaces on raised platforms that does not provide...
-
M303 Diagnostic Equipment Used by Patients in a Wheelchair (Section-by-Section Analysis)
M303 in the final rule establishes the technical requirements for diagnostic equipment used by patients seated in a wheelchair, such as weight scales with wheelchair spaces and mammography...
-
M303.1 General (Section-by-Section Analysis)
This is an introductory section.
-
M303.2 Wheelchair spaces (Section-by-Section Analysis)
This is an introductory section.
-
M303.2.1 Orientation (Section-by-Section Analysis)
The MDE NPRM proposed to require wheelchair spaces to be designed so that a patient in a wheelchair using diagnostic equipment would be oriented in the same direction that other non-wheelchair...
-
M303.2.2 Width (Section-by-Section Analysis)
The MDE NPRM proposed to require that diagnostic equipment used by patients seated in a wheelchair provide a wheelchair space that was at least 36 inches wide....
-
M303.2.3 Depth (Section-by-Section Analysis)
The MDE NPRM proposed two wheelchair space depth requirements based on how the wheelchair user enters the space: for spaces entered from the front or rear, 48 inches deep minimum; and for...
-
M303.2.4 Equipment Clearances (Section-by-Section Analysis)
The MDE NPRM proposed knee and toe clearance for diagnostic equipment used by patients seated in wheelchairs to allow for components in the wheelchair space which the patient could approach...
-
M303.2.5 Surfaces (Section-by-Section Analysis)
The MDE NPRM proposed to require diagnostic equipment used by patients seated in a wheelchair to provide a wheelchair space with a surface that does not slope more than 1:48 in any direction...
-
M303.2.6 Edge Protection (Section-by-Section Analysis)
(Significant Changes – Edge Protection), the final rule requires platforms with wheelchair spaces that are raised more than 1½ inches in height to provide a minimum 2-inch-high edge protection...
-
M303.3 Entry (Section-by-Section Analysis)
This is an introductory section.
-
M303.3.1 Vertical (Section-by-Section Analysis)
The MDE NPRM proposed that for equipment with a change in level at the entry to the wheelchair space, level changes of up to ¼ inch high are permitted to be vertical....
-
M303.3.2 Beveled (Section-by-Section Analysis)
The MDE NPRM proposed that for equipment with a change in level at the entry to the wheelchair space, level changes greater than ¼ inch but not greater than ½ inch would be required to be...
-
M303.3.3 Ramped (Section-by-Section Analysis)
The MDE NPRM proposed that for equipment with a change in level at the entry of a wheelchair space, level changes greater than ½ inch high would be required to be ramped and comply with...
-
M303.4 Components (Section-by-Section Analysis)
The MDE NPRM proposed to require diagnostic equipment used by patients seated in a wheelchair which has components that are used to examine specific body parts, be capable of examining those...
-
M303.4.1 Breast Platform Adjustability (Section-by-Section Analysis)
The MDE NPRM proposed a mammography breast platform height range of 30 inches high minimum and 42 inches high maximum above the floor. The Access Board received three comments on...
-
M303 Diagnostic Equipment Used by Patients Seated in a Wheelchair
-
M303 Diagnostic Equipment Used by Patients Seated in a Wheelchair
-
M303.1 General
Diagnostic equipment used by patients seated in a wheelchair shall comply with M303....
-
M303.1 General
Diagnostic equipment used by patients seated in a wheelchair shall comply with M303....
-
M303.2 Wheelchair Spaces
A wheelchair space complying with M303.2 shall be provided at diagnostic equipment. Advisory M303.2 Wheelchair Spaces....
-
M303.2 Wheelchair Spaces
Wheelchair spaces complying with M303.2 shall be provided at diagnostic equipment....