The existing guidelines include detailed requirements for the accessibility, usability, and compatibility of telecommunications equipment and customer premises equipment....
Search Results "Imaging Equipment with Bores"
Commonly Searched Documents
-
B. Proposed 255 Guidelines
-
M301.3.3 Head and Back Support
M301.3.3 Head and Back Support Proposed requirement: Where diagnostic equipment used by patients in a supine, prone, or side-lying position, and diagnostic equipment used by patients...
- UpScale by Medical Accessibility - Adjustable Height Exam Table with Built-in Scale
-
a. Knee and Toe Clearance
The requirements are intended to ensure that there is adequate space for a patient seated in a wheelchair to position underneath the equipment and align themselves against the breast platform...
-
M301.4 Lift Compatibility
Where the width of the examination surface is less than 36 inches (915 mm), the clearance depth shall extend the full width of the equipment....
-
11B-805.4.3 Turning space
A minimum of a 36-inch wide access shall be provided to and along each side of the bed, exam table, procedure table, gurney, lounge chair and/or equipment/apparatus. ...
- Living Earth Crafts Cloud 9 Cuvee Spa Treatment Table
- Living Earth Crafts Cloud 9 Spa Treatment Table
- Special Needs at Sea - Worldwide Equipment Rentals
-
7.3 Imaging System Accessibility Configuration Examples and Concerns
solutions to improving accessibility of existing imaging equipment and thus meet recommended standards for diagnostic equipment accessibility....
-
M302.4.2 Clearance Around Base
The base of the equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination surface...
-
1. Width and Depth of Wheelchair Spaces
The MDE NPRM proposed to require diagnostic equipment to have a wheelchair space that is at least 36 inches wide (proposed M303.2.2)....
-
M302.4.2 Clearance Around Base
The base of the diagnostic equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the examination...
-
M303.2.2 Width
Wheelchair spaces shall be 36 inches (915 mm) wide minimum. EXCEPTION: Wheelchair spaces located on raised platforms shall be permitted to be 32 inches (815 mm) wide minimum to a...
-
M303.3.1 Vertical
Changes in level of ¼ inch (6.4 mm) high maximum shall be permitted to be vertical.
-
M303.3.2 Beveled
Changes in level between ¼ inch (6.4 mm) high and ½ inch (13 mm) high maximum shall be beveled with a slope not steeper than 1:2.
-
M303.1 General (Section-by-Section Analysis)
This is an introductory section.
-
M303.2.4 Knee and Toe Clearance
Wheelchair spaces shall include knee and toe clearance complying with M303.2.4. The depth of the wheelchair space shall include knee and toe clearance of 17 inches (430 mm) minimum...
-
M303.3.1 Vertical
Changes in level of ¼ inch (6.4 mm) high maximum shall be permitted to be vertical.
-
M303.3.2 Beveled
Changes in level between ¼ inch (6.4 mm) high and ½ inch (13 mm) high maximum shall be beveled with a slope not steeper than 1:2.
-
Typical Transfer Techniques: Staff Assistance and Patient Lifts
The provider should ask the patient if he or she needs assistance, and if so, what is the best way to help and what extra equipment, if any, is needed....
- Midmark 623 Barrier-Free® Power Examination Table
- Midmark 625 Barrier-Free® Power Examination Table
- Midmark Ritter 222 Barrier-Free® Power Examination Table