A transfer surface shall be provided and shall comply with M301.2.
Search Results "Adjustable Height Counter"
-
M301.2 Transfer Surface
-
M301.1 General
Diagnostic equipment that supports patients in a supine, prone, or side-lying position shall comply with M301. EXCEPTION: Examination chairs complying with M302 that recline to...
-
M301.2.3 Size
The size of the transfer surface shall comply with M301.2.3.1 or M301.2.3.2. The size of transfer surfaces shall be measured from center points of their opposing sides.
-
M301.2.3 Size (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface size for diagnostic equipment used in the supine, prone, or side-lying position of 30 inches wide and 15 inches deep minimum. (proposed...
-
5.2.1 Transfer Surface Size Recommendations for M301
5.2.1 Transfer Surface Size Recommendations for M301 Description: The transfer surface is the part of the diagnostic equipment onto which patients who use mobility devices or aids...
-
5.2.1.1 Transfer Surface Width
5.2.1.1 Transfer Surface Width. The transfer surface should be 28 inches wide minimum for equipment used by patients in a supine, side-lying, or prone position.
-
5.2 Transfer Surface Size for Equipment Used by Patients in Supine, Prone, or Side-Lying Position
5.2 Transfer Surface Size for Equipment Used by Patients in Supine, Prone, or Side-Lying Position.
-
5.2.1 Transfer Surface Size
5.2.1 Transfer Surface Size
-
5.2.1.2 Transfer Surface Depth
5.2.1.2 Transfer Surface Depth. The transfer surface should be 17 inches deep minimum for all equipment used by patients in a supine, side-lying, or prone position with the exception...
-
5.2 Transfer Surface Size Recommendations
5.2 Transfer Surface Size Recommendations
-
Examples of access:
Exam table that lowers to 17-19 inches from the floor Many people find these tables easier to use. There are many ways to help people get access. See the [images]...
-
M301.4.1 Clearance in Base
[See subsections ...]
-
M301.2.2 Size
The transfer surface shall be 30 inches (760 mm) wide minimum and 15 inches (381mm) deep minimum. Advisory M301.2.2 Size. The size requirements in this section apply only to the...
-
M301.2 Transfer Surface
A transfer surface shall be provided and shall comply with M301.2.
-
M301.1 General
Where diagnostic equipment is used by patients in a supine, prone, or side-lying position, it shall comply with M301.
-
602.4 Spout Height
Spout outlets shall be 36 inches (915 mm) maximum above the finish floor or ground.
-
602.4 Spout Height
Spout outlets shall be 36 inches (915 mm) maximum above the finish floor or ground.
-
4.15.2* Spout Height
Spouts shall be no higher than 36 in (915 mm), measured from the floor or ground surfaces to the spout outlet (see Fig. 27(a)).
-
4.15.2* SPOUT HEIGHT
Spouts shall be no higher than 36 in (915 mm), measured from the floor or ground surfaces to the spout outlet (see Fig. 27(a)). Figure 27 Drinking Fountains and Water...
-
Height: ADA Standard Section 704.4.1
Advisory 704.4.1 Height....
-
Height: ADA Standard Section 1009.5.3
The height of the transfer platform shall comply with 1009.4.2....
-
Adjustable height protocol
Adjustable height protocol: From initial setup, only the height of the platform was adjusted incrementally: higher (Figure 3) and lower than the subject’s seat. ...
-
Kitchens, Kitchenettes, or Wet Bars [9.2.2(7)]
Kitchens, Kitchenettes, or Wet Bars [9.2.2(7)] Clear floor space for either a forward or a side approach is permitted for cabinets, counters, sinks, and appliances....
-
Height: ADA Standard Section 1009.4.2
The height of the transfer wall shall be 16 inches (405 mm) minimum and 19 inches (485 mm) maximum measured from the deck....