5.3.2 Transfer Sides Recommendations for M301
-
M301.3.1 Transfer Supports
Transfer supports shall be provided for use with the transfer sides required by M301.2.3 and shall comply with M305.2.
-
M301.3 Supports
Transfer supports, stirrups, and reclining surfaces shall comply with M301.3.
-
M304.2.2 Standing Supports (Section-by-Section Analysis)
The MDE NPRM proposed requiring standing supports on each side of the standing surface of diagnostic equipment used by patients in the standing position, and compliance with the technical...
-
B. Legal foundation for captioning and video description
The statute defines auxiliary aids to include "qualified interpreters or other effective methods of making aurally delivered materials available to individuals with hearing impairments"...
-
2. What are examples of auxiliary aids and services for students with hearing, vision, and speech disabilities?
In general, auxiliary aids and services make aurally or visually delivered information available to students with hearing, vision, or speech disabilities so that they can receive information...
-
4.2 Functional Comparisons
Unless more hard of hearing people can be convinced (or their hearing aid dispensers) to include telecoils in their hearing aids, this situation is unlikely to change....
-
2.3.2 IR Receivers
Some under-chin receivers include an output jack into which a neckloop can be plugged for inductive coupling to a hearing aid....
-
M304.2.2 Standing Supports
be permitted to provide one standing support on one side of the standing surface provided that the standing support complies with the requirements for standing supports in a horizontal position...
-
Figure M4: Overview of M303 Diagnostic Equipment for Use by a Patient While Seated in a Wheelchair and M304 Diagnostic Equipment for Use by a Patient in a Standing Position (Weight Scale Example)
Where the primary use of the equipment supports more than one patient position, it must comply with the requirements for each use. ...
-
Attachment C
Various types of communication aids – known as “auxiliary aids and services” – are used to communicate with people who are deaf or hard of hearing. ...
-
3.Guidelines for Compatibility With Peripheral Devices
be compatible with peripheral devices commonly used by individuals with disabilities to achieve accessibility, and should comply with the following provisions, as applicable: Hearing aid...
-
d. Unobstructed Transfer
The final rule retains the proposed requirements for unobstructed transfer for diagnostic equipment used in a supine, prone, or side-lying position, as well as diagnostic equipment used...
-
M301.2.3.1 End Transfer Surface
End transfer surfaces shall be 28 inches (710 mm) wide minimum and 17 inches (430 mm) long minimum. EXCEPTION: Transfer surfaces for imaging equipment with bores shall be...
-
M301.2.3.2 Side Transfer Surface
Side transfer surfaces shall be 28 inches (710 mm) wide minimum and 28 inches (710 mm) long minimum. EXCEPTION: Transfer surfaces for imaging equipment with bores shall be...
-
M301.2.4 Unobstructed Transfer
Each transfer surface shall provide two unobstructed sides for patient transfer. EXCEPTIONS: 1. Obstructions no more than 3 inches (75 mm) deep shall be permitted to extend beyond...
-
M301.2.4 Unobstructed Transfer (Section-by-Section Analysis)
In the MDE NPRM the Access Board proposed that each transfer side provide unobstructed access to the transfer surface, with an exception to permit temporary obstructions as long as...
-
M301.2.3 Transfer Sides
The transfer surface shall be located to provide options to transfer from a mobility device onto one short side (depth) and one long side (width) of the surface. Each transfer side...
-
410.3 Magnetic Coupling
delivers output by an audio transducer that is typically held up to the ear, ICT shall provide a means for effective magnetic wireless coupling to hearing technologies, such as hearing aids...
-
III-4.6100 Examinations
Where necessary, an examiner may be required to provide auxiliary aids or services, unless it can demonstrate that offering a particular auxiliary aid or service would fundamentally alter...
-
11B-608.6 Shower spray unit and water
A shower spray unit with a hose 59 inches (1499 mm) long minimum that can be used both as a fixed-position shower head and as a hand-held shower shall be provided....
-
11B-608.6 Shower spray unit and water
A shower spray unit with a hose 59 inches (1499 mm) long minimum that can be used both as a fixed-position shower head and as a hand-held shower shall be provided....
-
5.2.3 Transfer Surface Measurement Recommendation for M301 and M302
The transfer surface dimensions do not include headrests, footrests, or similar supports for body extremities that do not support the patient’s overall body position....