The size of the transfer surface shall comply with M301.2.3.1 or M301.2.3.2. The size of transfer surfaces shall be measured from center points of their opposing sides.
Search Results "Treatment Table"
-
M301.2.3 Size
-
36 CFR Part 1195 Standards for Accessible Medical Diagnostic Equipment
The standards for medical diagnostic equipment (MDE Standards) contain minimum technical criteria to ensure that medical diagnostic equipment, including but not limited to, examination tables...
-
36 CFR Part 1195 Standards for Accessible Medical Diagnostic Equipment - Preamble
The standards for medical diagnostic equipment (MDE Standards) contain minimum technical criteria to ensure that medical diagnostic equipment, including but not limited to, examination tables...
- Kompan, Inc. M7000 - Red House Playhouse
-
11B-604.8.1.1 Size
This space cannot be obstructed by baby changing tables or other fixtures or conveniences, except as specified at 604.3.2 (Overlap)....
-
M301.3 Supports (Section-by-Section Analysis)
This is an introductory section. An editorial change was made to this section as a result of the change in M301.3.2, described below, to replace the word “stirrups” with the term...
-
M301.3.3 Head and Back Support
Where the diagnostic equipment is used in a reclined position, head and back support shall be provided. Where the incline of the back support can be modified while in use, head and...
-
Table 2: Summary of Department of Transportation Survey
Architectural Research Consulting 20-07 (167) No. Element surveyed Minimum reported tolerance Maximum reported tolerance Average reported...
-
11B-603.5 Accessories
Where towel or sanitary napkin dispensers, waste receptacles, or other accessories are provided in toilet facilities, at least one of each type shall be located on an accessible...
-
M301.3.2 Leg Supports
Where stirrups are provided, leg supports shall also be provided and shall comply with M305.4.
-
5.2.1.1 Transfer Surface Width
5.2.1.1 Transfer Surface Width. The transfer surface should be 28 inches wide minimum for equipment used by patients in a supine, side-lying, or prone position.
-
M301.4.1 Clearance in Base
-
5.2.1.2 Transfer Surface Depth
5.2.1.2 Transfer Surface Depth. The transfer surface should be 17 inches deep minimum for all equipment used by patients in a supine, side-lying, or prone position with the exception...
-
5.2 Transfer Surface Size Recommendations
5.2 Transfer Surface Size Recommendations
-
M301.3.2 Leg Supports (Section-by-Section Analysis)
In the MDE NPRM, the Access Board proposed to place the requirements for stirrups on diagnostic equipment used by patients in the supine, prone, or side-lying position in M301. In...
-
M301.3.2 Stirrups
Where stirrups are provided, they shall provide a method of supporting, positioning, and securing the patient’s legs.
-
M301.3.1 Transfer Supports
Transfer supports shall be provided for use with the transfer sides required by M301.2.3 and shall comply with M305.2.
-
M301.3 Supports
Transfer supports, leg supports, and reclining surfaces shall comply with M301.3.
-
5.3.2 Transfer Sides for Equipment Used by Patients in a Supine, Prone, or Side-Lying Position
5.3.2 Transfer Sides for Equipment Used by Patients in a Supine, Prone, or Side-Lying Position.
- Figure 604.8.1.1 Size of Wheelchair Accessible Toilet Compartment – Drawing from the U.S. Access Board
-
M305.2.2 Length (Section-by-Section Analysis)
imaging equipment with transfer surfaces less than or equal to 24 inches deep, the Committee recommended requiring a transfer support to extend horizontally along the side of the patient table...
-
M301.4 Lift Compatibility
Diagnostic equipment shall be usable with a patient lift and shall comply with M301.4.1 or M301.4.2.
-
M301.2.3.1 End Transfer Surface
End transfer surfaces shall be 28 inches (710 mm) wide minimum and 17 inches (430 mm) long minimum. EXCEPTION: Transfer surfaces for imaging equipment with bores shall be...
-
M301.2.4 Unobstructed Transfer
Each transfer surface shall provide two unobstructed sides for patient transfer. EXCEPTIONS: 1. Obstructions no more than 3 inches (75 mm) deep shall be permitted to extend beyond...