Transfer supports shall be located within reach of the transfer surface and shall not obstruct transfer onto or off of the surface when in position.
Search Results "Table"
Commonly Searched Documents
-
M305.2.1 Location
-
M301.2.1 Adjustability
Transfer surfaces shall be adjustable in height measured from the floor to the top of the uncompressed transfer surface and shall provide the following: A. A low transfer position...
-
M301.2.2 Sunset
The low transfer position height, Item A of M301.2.1, shall cease to have effect on [Insert date 5 years after date of publication in the Federal Register].
-
d. Unobstructed Transfer
The MDE NPRM proposed that each transfer side provide unobstructed access to the transfer surface, with an exception to permit temporary obstructions as long as they could be...
-
M301.2.1 Adjustability (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface height range for diagnostic equipment used by patients in the supine, prone, or side-lying position of 17 inches minimum and 19 inches...
-
M301.2.2 Sunset Provision (Section-by-Section Analysis)
As discussed in Section IV.B.1.a. (Significant Changes – Transfer Surface Adjustability), this is a new provision that was added to the final rule in conjunction with the new...
-
M301.4 Lift Compatibility (Section-by-Section Analysis)
The MDE NPRM proposed to require that diagnostic equipment used by patients in the supine, prone, or side-lying position be usable with a patient lift and comply with either the...
-
5.4.7 Transfer Support Gripping Surface Clearance Recommendation for M301 and M302
5.4.7 Transfer Support Gripping Surface Clearance Recommendation for M301 and M302 Description: A clearance around the gripping surface is necessary to ensure sufficient space for...
-
5.7 Lift Compatibility
5.7 Lift Compatibility.
-
5.4.7 Transfer Support Gripping Surface Clearance for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.7 Transfer Support Gripping Surface Clearance for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position. There should be a 1½ inches minimum clearance around the...
-
5.7 Lift Compatibility Recommendations
5.7 Lift Compatibility Recommendations
-
5.4.4 Transfer Support Distance From Transfer Surface for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.4 Transfer Support Distance From Transfer Surface for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position. The transfer support should be located a maximum distance...
-
(2) Adjustability: Minimum Low Transfer Height
In the 11 years we have been selling adjustable height examination tables we do not have a single complaint on record regarding the accessibility of our 19” low height tables....
-
M305.2.5 Surface Hazards
Transfer supports and surfaces adjacent to transfer supports shall be free of sharp or abrasive components and shall have eased edges.
-
M305.2 Transfer Supports
Transfer supports shall comply with M305.2.
-
M305.2 Transfer Supports (Section-by-Section Analysis)
This is an introductory section. As discussed above in Section IV.E.1. (Significant Changes – Transfer Supports), the Access Board strengthened the transfer support requirements and...
-
M305.2.5 Surface Hazards (Section-by-Section Analysis)
The proposed rule did not provide any specific restrictions regarding surface hazards around the transfer supports. No public comments were submitted on this issue, but the MDE...
-
5.4.6 Transfer Support Gripping Surface Cross Section and Clearances for M301, M302, and M305.2
5.4.6 Transfer Support Gripping Surface Cross Section and Clearances for M301, M302, and M305.2 Description: This provision prescribes a cross-sectional profile of the transfer...
-
5.4 Transfer Supports
5.4 Transfer Supports.
-
5.4 Transfer Supports Recommendations
5.4 Transfer Supports Recommendations
-
M305.2 Transfer Supports
Transfer supports shall comply with M305.2.
-
4.1.6 Surface
The surface of the clear floor or ground space shall be firm and stable. The type of surface should be appropriate to the setting and level of development.
-
M305.2.2.1 End Transfer Supports
Transfer supports for transfer surfaces complying with M301.2.3.1 and M305.2.2.1 shall be 15 inches (380 mm) long minimum. Transfer supports shall be positioned along 13 ½ inches...