Transfer surfaces shall be adjustable in height measured from the floor to the top of the uncompressed transfer surface and shall provide the following: A. A low transfer position...
Search Results "Table"
Commonly Searched Documents
-
M301.2.1 Adjustability
-
M301.2.2 Sunset
The low transfer position height, Item A of M301.2.1, shall cease to have effect on [Insert date 5 years after date of publication in the Federal Register].
-
d. Unobstructed Transfer
The MDE NPRM proposed that each transfer side provide unobstructed access to the transfer surface, with an exception to permit temporary obstructions as long as they could be...
-
M301.2.1 Adjustability (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface height range for diagnostic equipment used by patients in the supine, prone, or side-lying position of 17 inches minimum and 19 inches...
-
M301.2.2 Sunset Provision (Section-by-Section Analysis)
As discussed in Section IV.B.1.a. (Significant Changes – Transfer Surface Adjustability), this is a new provision that was added to the final rule in conjunction with the new...
-
M301.4 Lift Compatibility (Section-by-Section Analysis)
The MDE NPRM proposed to require that diagnostic equipment used by patients in the supine, prone, or side-lying position be usable with a patient lift and comply with either the...
-
5.4.7 Transfer Support Gripping Surface Clearance Recommendation for M301 and M302
5.4.7 Transfer Support Gripping Surface Clearance Recommendation for M301 and M302 Description: A clearance around the gripping surface is necessary to ensure sufficient space for...
-
5.7 Lift Compatibility
5.7 Lift Compatibility.
-
5.4.7 Transfer Support Gripping Surface Clearance for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.7 Transfer Support Gripping Surface Clearance for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position. There should be a 1½ inches minimum clearance around the...
-
5.7 Lift Compatibility Recommendations
5.7 Lift Compatibility Recommendations
-
5.4.4 Transfer Support Distance From Transfer Surface for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.4 Transfer Support Distance From Transfer Surface for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position. The transfer support should be located a maximum distance...
-
(2) Adjustability: Minimum Low Transfer Height
In the 11 years we have been selling adjustable height examination tables we do not have a single complaint on record regarding the accessibility of our 19” low height tables....
-
M305.2.5 Surface Hazards
Transfer supports and surfaces adjacent to transfer supports shall be free of sharp or abrasive components and shall have eased edges.
-
M305.2 Transfer Supports
Transfer supports shall comply with M305.2.
-
M305.2 Transfer Supports (Section-by-Section Analysis)
This is an introductory section. As discussed above in Section IV.E.1. (Significant Changes – Transfer Supports), the Access Board strengthened the transfer support requirements and...
-
M305.2.5 Surface Hazards (Section-by-Section Analysis)
The proposed rule did not provide any specific restrictions regarding surface hazards around the transfer supports. No public comments were submitted on this issue, but the MDE...
-
5.4.6 Transfer Support Gripping Surface Cross Section and Clearances for M301, M302, and M305.2
5.4.6 Transfer Support Gripping Surface Cross Section and Clearances for M301, M302, and M305.2 Description: This provision prescribes a cross-sectional profile of the transfer...
-
5.4 Transfer Supports
5.4 Transfer Supports.
-
5.4 Transfer Supports Recommendations
5.4 Transfer Supports Recommendations
-
M305.2 Transfer Supports
Transfer supports shall comply with M305.2.
-
4.1.6 Surface
The surface of the clear floor or ground space shall be firm and stable. The type of surface should be appropriate to the setting and level of development.
-
M305.2.2.1 End Transfer Supports
Transfer supports for transfer surfaces complying with M301.2.3.1 and M305.2.2.1 shall be 15 inches (380 mm) long minimum. Transfer supports shall be positioned along 13 ½ inches...
-
M305.2.4.1 Circular Cross Section
Transfer supports with circular cross sections shall have an outside diameter of 1 ¼ inches (32 mm) minimum and 2 inches (51 mm) maximum.