The transfer surface should be 28 inches wide minimum for equipment used by patients in a supine, side-lying, or prone position....
Search Results "Positioning Support"
Commonly Searched Documents
-
5.2.1.1 Transfer Surface Width
-
5.2.1.2 Transfer Surface Depth
The transfer surface should be 17 inches deep minimum for all equipment used by patients in a supine, side-lying, or prone position with the exception of imaging equipment....
-
5.2.1.4 Transfer Surface Size for Imaging Equipment
5.2.1.4 Transfer Surface Size for Imaging Equipment. The transfer surface should be 28 inches wide minimum and 21 inches deep minimum. The transfer surface should be located such...
-
(2) Transfer Surface Location for Diagnostic Equipment Used in the Seated Position
Commenters also raised concerns with the provisions in the MDE NPRM related to transferring to medical diagnostic equipment used by patients in the seated position....
-
11B-411.4.7 Designations and indicators of car control buttons
Designations and indicators of car control buttons shall comply with Section 11B-411.4.7.
-
c. Transfer Surface Size for Diagnostic Equipment Used by Patients in the Supine, Prone, or Side-lying Position
Of the 12 commenters who responded, only two supported the transfer surface size in the proposed rule....
-
Shower Spray Unit
The guidelines require tubs to have shower spray units that can be used as both a fixed position shower head and a hand-held shower (607.6)....
-
Implication for Standards and Design
Each phase involved a highly customized transfer station iteratively developed with scientific, expert and consumer input to support the goals of the project. ...
-
M301.2.1 Adjustability (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface height range for diagnostic equipment used by patients in the supine, prone, or side-lying position of 17 inches minimum and 19 inches maximum....
-
M302.1 General
Where diagnostic equipment is used by patients in a seated position, it shall comply with M302....
-
5.2.1 Transfer Surface Size Recommendations for M301
The technical criteria do not address the overall width and depth of patient support surfaces because of the diverse shape and size of these surfaces....
-
5.2.2.2 Transfer Surface Depth Recommendation for M302
5.2.2.2 Transfer Surface Depth Recommendation for M302 The Committee recommends a minimum depth requirement of 17 inches for equipment used by patients in seated positions....
-
M301.2.3 Size (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface size for diagnostic equipment used in the supine, prone, or side-lying position of 30 inches wide and 15 inches deep minimum....
-
5.2.1.1 Transfer Surface Width Recommendations for M301
Transfer Surface Width Recommendations for M301 The Committee recommends a transfer surface of 28 inches minimum width for equipment used by patients in a supine, side-lying, or prone position...
-
5.1.2 Transfer Surface High Height Recommendations for M301 and M302
5.1.2 Transfer Surface High Height Recommendations for M301 and M302 The Committee recommends an adjustable height range with an upper height measuring at least 25...
-
5.1.3 Transfer Surface Low Height Options for M301 and M302
5.1.3 Transfer Surface Low Height Options for M301 and M302 The Committee did not agree on a final low height recommendation. Rationale for final recommendation (more detail...
-
5.1.3 Transfer Surface Low Height
5.1.3 Transfer Surface Low Height. No agreement on final low height
-
5.1.2 Transfer Surface High Height
5.1.2 Transfer Surface High Height. The transfer surface should have an adjustable height range with an upper height measuring at least 25 inches.
-
5.1.1 Transfer Surface Height Adjustability
5.1.1 Transfer Surface Height Adjustability. The transfer surface should have an adjustable height in small virtually continuous increments.
-
5.3.2.2 Transfer Sides for Imaging Equipment
5.3.2.2 Transfer Sides for Imaging Equipment. The transfer surface should be provided on at least one long side of the table.
- Service Animals and Emotional Support Animals
-
M302.2.1 Adjustability
A low transfer position at a height of 17 inches (430 mm) minimum and 19 inches (485 mm) maximum; B. A high transfer position at 25 inches (635 mm); and C....
-
M302.2.1 Adjustability (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface height range for diagnostic equipment used by patients in a seated position of 17 inches minimum and 19 inches maximum....
-
M302.2.3 Size (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface size for diagnostic equipment used by patients in the seated position of 21 inches wide and 15 inches deep (proposed M302.2.2)....