5.8.2 Knee and Toe Clearance Under Breast Platform....
Search Results "Breast Platform"
-
5.8.2 Knee and Toe Clearance Under Breast Platform
-
5.8.2.1 Height of Breast Platform At Time of Measurement
5.8.2.1 Height of Breast Platform At Time of Measurement. All knee and toe clearances should be measured when the top of the breast platform is set at 34 inches above the floor....
-
5.8.2.2 Knee Clearance Depth at 27” Above the Ground
5.8.2.2 Knee Clearance Depth at 27” Above the Ground. The knee clearance should be increased at 27 inches above the ground to 18 inches minimum.
-
5.8.2.3 Overall Knee and Toe Clearance
5.8.2.3 Overall Knee and Toe Clearance. The overall knee and toe space should be increased to a minimum of 28 inches.
-
5.8.2.4 Unobstructed Floor Space
5.8.2.4 Unobstructed Floor Space. The unobstructed floor space in front of the base support should be a minimum of 17 inches deep.
-
5.8.2.5 Toe Height
The toe height should be 18 inches minimum, measured above the ground when the top of the breast platform is 34 inches above the ground....
-
5.8.2.6 Clearance Depth at Toe Height Above the Ground
5.8.2.6 Clearance Depth at Toe Height Above the Ground. The toe clearance depth should be 22 inches minimum at toe height.
-
5.8.2.7 Base Support Allowance
An additional sloped region above the base support is permitted at a depth of 25 inches from the front edge of the breast platform at 1½ inches above the floor, and can extend to a height...
-
5.8.3 Breast Platform
5.8.3 Breast Platform...
-
5.8.3.1 Breast Platform Height
5.8.3.1 Breast Platform Height. The height of the top of the breast platform should be decreased to 26 inches above the floor....
-
2. Equipment Clearances for Breast Platforms
a breast platform would be 25 inches deep (proposed M303.2.4)....
-
a. Knee and Toe Clearance
The proposed rule recommended a knee and toe clearance depth for breast platforms of 25 inches. There were no comments received on this requirement....
-
b. Exception for Base Support Allowance and Unobstructed Knee and Toe Space
In the final rule, the Access Board has created an exception to the height requirement for breast platforms....
-
4. Breast Platform Adjustability
Additionally, the Access Board proposed specific technical requirements for breast platforms of mammography equipment....
-
1. Standing Supports
The second commenter, a manufacturer, expressed concerns that the proposed rule was treating supports on breast platforms as standing supports, explaining that this was not the supports’...
-
M303.2.4 Equipment Clearances (Section-by-Section Analysis)
The proposed rule provided one additional requirement for breast platforms on mammography equipment, proposing the knee and toe clearance under a breast platform to be 25 inches deep (proposed...
-
M303.4 Components (Section-by-Section Analysis)
For example, an x-ray platform on which a patient places an arm or hand would have to be capable of examining the arm or hand of the patient while seated in a wheelchair....
-
M303.4.1 Breast Platform Adjustability (Section-by-Section Analysis)
The MDE NPRM proposed a mammography breast platform height range of 30 inches high minimum and 42 inches high maximum above the floor....
-
M303.2.4.1 Breast Platforms
Wheelchair spaces beneath breast platforms shall comply with M303.2.4.1....
-
M303.2.4.1.1 Depth
Wheelchair spaces shall include knee and toe clearance 25 inches (635 mm) deep minimum and 28 inches (710 mm) deep maximum.
-
M303.2.4.1.2 Height
22 inches (560 mm) maximum at 9 inches (230 mm) above the floor and at a depth of 16 inches (405 mm) minimum at 27 inches (685 mm) above the floor measured from the leading edge of the breast...
-
M303.4 Components
Breast platforms shall comply with M303.4.1....
-
M303.4 Components
Breast platforms shall comply with M303.4.1....
-
M303.4.1 Breast Platforms
The height of the breast platform shall be 30 inches (760 mm) high minimum and 42 inches (1065 mm) high maximum above the floor when in use by a patient seated in a wheelchair....