M303.1 General
Diagnostic equipment used by patients seated in a wheelchair shall comply with M303.
-
M303.2.2 Width
Wheelchair spaces shall be 36 inches (915 mm) wide minimum. EXCEPTION: Wheelchair spaces located on raised platforms shall be permitted to be 32 inches (815 mm) wide minimum to a...
-
M303.3.1 Vertical
Changes in level of ¼ inch (6.4 mm) high maximum shall be permitted to be vertical.
-
M303.3.2 Beveled
Changes in level between ¼ inch (6.4 mm) high and ½ inch (13 mm) high maximum shall be beveled with a slope not steeper than 1:2.
-
M303.4 Components
Where components of diagnostic equipment are used to examine specific body parts, the components shall be capable of examining the body parts of a patient seated in a wheelchair....
-
M303.3.2 Beveled (Section-by-Section Analysis)
The MDE NPRM proposed that for equipment with a change in level at the entry to the wheelchair space, level changes greater than ¼ inch but not greater than ½ inch would be required...
-
M303.2.2 Width (Section-by-Section Analysis)
The MDE NPRM proposed to require that diagnostic equipment used by patients seated in a wheelchair provide a wheelchair space that was at least 36 inches wide. There were no public...
-
M303.3.1 Vertical (Section-by-Section Analysis)
The MDE NPRM proposed that for equipment with a change in level at the entry to the wheelchair space, level changes of up to ¼ inch high are permitted to be vertical. This provision...
-
M303.4 Components (Section-by-Section Analysis)
The MDE NPRM proposed to require diagnostic equipment used by patients seated in a wheelchair which has components that are used to examine specific body parts, be capable of...
-
M303.1 General (Section-by-Section Analysis)
This is an introductory section.
-
M303.4 Components
Where components of diagnostic equipment are used to examine specific body parts, the components shall be capable of examining the body parts of a patient seated in a wheelchair....
-
M303.3.1 Vertical
Changes in level of ¼ inch (6.4 mm) high maximum shall be permitted to be vertical.
-
M303.3.2 Beveled
Changes in level between ¼ inch (6.4 mm) high and ½ inch (13 mm) high maximum shall be beveled with a slope not steeper than 1:2.
-
M303.1 General
Diagnostic equipment used by patients seated in a wheelchair shall comply with M303.
-
A. Medical equipment and furniture
Medical equipment and furniture Without accessible medical examination tables, dental chairs, radiological diagnostic equipment, scales, and rehabilitation equipment, individuals with...
- UpScale by Medical Accessibility - Adjustable Height Exam Table with Built-in Scale
-
Ramps M303:3.3, 3.3.1, 3.3.2
A ramp, using a 1:2 slope, will allow for short rise ramps found on stow-a-weigh scales....
-
Line of Sight and Dispersion of Wheelchair Spaces in Assembly Areas
sight in the 1991 Standards to mean wheelchair spaces and companion seats in sports stadia and arenas must provide patrons with disabilities and their companions with lines of sight over standing...
-
M303.3.3.3 Clear Width
The clear width of ramp runs shall be 36 inches (915 mm) minimum.
-
M303.3.3.1 Running Slope
Ramp runs shall have a running slope not steeper than 1:12. EXCEPTION: A running slope not steeper than 1:8 shall be permitted for ramp runs with a maximum height of 2 ½ inches...
-
M303.3.3.4 Edge Protection
Ramps with drop offs ½ inch (13 mm) or greater shall provide edge protection 2 inches (50 mm) high minimum on each side with a drop off.