The transfer surface shall be 30 inches (760 mm) wide minimum and 15 inches (381mm) deep minimum. Advisory M301.2.2 Size. The size requirements in this section apply only to the...
Search Results "Patient in Side-Lying Position"
Commonly Searched Documents
-
M301.2.2 Size
-
M301.2 Transfer Surface
A transfer surface shall be provided and shall comply with M301.2.
-
M301.3.1 Transfer Supports
Transfer supports shall be provided for use with the transfer sides required by M301.2.3 and shall comply with M305.2....
-
M305.2.6 Gripping Surface
Transfer support gripping surfaces shall be continuous along their length and shall not be obstructed along their tops or sides....
-
M305.5 Head and Back Support
Where the diagnostic equipment is used in a reclined position, head and back support shall be provided....
-
M301.3.3 Head and Back Support
Where the diagnostic equipment is used in a reclined position, head and back support shall be provided....
-
M305.2.2 Length
The length of transfer supports shall comply with M305.2.2.1 or M305.2.2.2.
-
5.1.1 Transfer Surface Height Adjustability Recommendation for M301 and M302
Adjustable height MDE, such as exam tables, imaging tables and chairs, will make it possible to position the transfer surface near the height of the seat of the mobility device....
-
M302.2.4 Transfer Sides (Section-by-Section Analysis)
In the MDE NPRM, the transfer side provision for diagnostic equipment used by patients in the seated position required transfer surfaces to have the option to transfer from a mobility device...
-
M301.2.3.1 End Transfer Surface
imaging equipment with bores shall be permitted to be 21 inches (535 mm) wide minimum but shall not be permitted to be less than the full width of the examination surface provided for the patient...
-
5.7.1 Lift Compatibility Clearance in Base for M301
portable floor lifts to be effective, they must deploy at diagnostic equipment such that the boom of the lifts can maneuver far enough over the equipment to safely lower and raise the patient...
-
5.1.2 Transfer Surface High Height Recommendations for M301 and M302
Therefore, transfer surfaces that are adjustable to a 25 inches maximum during patient transfer accommodate most patients who use mobility devices....
-
5.4.8 Transfer Support Gripping Surface Hazard Recommendation for M301 and M302
Rationale for the recommendation Gripping surface configurations must provide an effective and safe surface for patients to hold onto....
-
M301.3.2 Leg Supports
Where stirrups are provided, leg supports shall also be provided and shall comply with M305.4.
-
M305.2.8 Fittings
Transfer supports shall not rotate within their fittings when in place for transfer.
-
M305.2.7 Clearance
Clearance between the transfer support gripping surface and adjacent surfaces or obstructions shall be 1 ½ inches (38 mm) minimum.
-
M305.2.7 Clearance (Section-by-Section Analysis)
In the MDE NPRM, the Access Board did not provide any specific requirements for clearances around the transfer support. However, in the preamble to the MDE NPRM the Access Board...
-
M305.2.3 Fittings
Transfer supports shall not rotate within their fittings.
-
M305.2.1 Location
Transfer supports shall comply with M305.2.1.1 or M305.2.1.2 and shall be located 1 ½ inches (38 mm) maximum measured horizontally from the plane defined by the nearest edge of the...
-
M305.2.4 Cross Section
Transfer supports shall have a cross section complying with 305.2.4.1 or 305.2.4.2.
-
M305.2.3 Height
During use, the tops of transfer support gripping surfaces shall be 6 inches (150 mm) minimum and 19 inches (485 mm) maximum higher than the top of the associated uncompressed...
-
M305.2.4 Cross Section (Section-by-Section Analysis)
The proposed rule did not provide specific requirements for the cross section of transfer supports. However, in the MDE NPRM preamble, the Access Board noted that it was considering...
-
M305.2.1.1 End Transfer Supports
Transfer supports for transfer surfaces complying with M301.2.3.1 and M302.2 shall be located on the short side (length) opposite the transfer side....
-
M301.3.1 Transfer Supports
Transfer surfaces required by M301.2 shall provide transfer supports and shall comply with M305.2.