Signs shall comply with Section 11B-703. Where both visual and tactile characters are required, either one sign with both visual and tactile characters, or two separate signs, one...
Search Results "Seat Designation"
-
11B-703.1 General
-
What types of facilities are covered by DOT’s ADA Standards (and are required to have detectable warnings on curb ramps)?
DOT’s ADA Standards apply to facilities used to provide “designated public transportation.”...
-
Whitehall Mfg. WH2125A On-Floor ADA Compliant Bariatric Stainless Steel Toilet
Requires the Premium User plan or above. (Login or upgrade your account!)
- Figure 608.3.2 Grab Bars for Standard Roll-In Type Showers – Drawing from the U.S. Access Board
- The RERC on Workplace Accommodations
-
4. Exception from the Requirements of M301 for Certain Examination Chairs that Comply with M302
The Access Board proposed in M101.2 in the MDE NPRM to require diagnostic equipment to meet the standards for each patient position supported, meaning that if diagnostic equipment was designed...
-
4.4.4 Clear Floor or Ground Space
Locate the clear space to provide shoulder alignment between a person sitting on the bench and a person seated in a wheelchair occupying the clear space....
- Access Board: Tactile Sign Braille Requirements
-
Plan 6A: 12-Foot Wide Accessible Guest Room
Furnishings include a king bed and additional seating. ...
-
Plan 2A: 13-Foot Wide Accessible Guest Room
Features include a standard roll-in shower with a seat, comparable vanity, wardrobe, and door connecting to adjacent guest room....
-
5.4.9 Interruptions Along Transfer Support Gripping for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.9 Interruptions Along Transfer Support Gripping for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position....
-
Plan 4A: 13-Foot Wide Accessible Guest Room
Features include a standard roll-in shower with a seat, comparable vanity, clothes closet with swinging doors, and door connecting to adjacent guest room....
-
M303.3.3.3 Clear Width
The clear width of ramp runs shall be 36 inches (915 mm) minimum.
-
M303.2.4.2 Other Equipment
Wheelchair spaces beneath diagnostic equipment other than breast platforms shall comply with M303.2.4.2.
-
M302.2.5 Unobstructed Transfer
Each transfer side complying with M302.2.4 shall provide unobstructed access to transfer surfaces. EXCEPTIONS: 1. Obstructions no more than 3 inches (75 mm) deep shall be...
-
M303.3.3.1 Running Slope
Ramp runs shall have a running slope not steeper than 1:12. EXCEPTION: A running slope not steeper than 1:8 shall be permitted for ramp runs with a maximum height of 2 ½ inches...
-
M303.3.3.4 Edge Protection
Ramps with drop offs ½ inch (13 mm) or greater shall provide edge protection 2 inches (50 mm) high minimum on each side with a drop off.
-
M303.3.3.5 Handrails
Ramps with a rise greater than 6 inches (150 mm) shall provide handrails on both sides.
-
M303.2.6 Edge Protection (Section-by-Section Analysis)
The MDE NPRM proposed edge protection on the ramps leading up to the raised platform (proposed M303.3.3.4), but did not require edge protection on the raised platforms themselves....
-
M302.2.5 Unobstructed Transfer (Section-by-Section Analysis)
In the MDE NPRM the Access Board proposed that each transfer side provide unobstructed access to the transfer surface, with an exception to permit temporary obstructions as long as...
-
5.8.1.1 Clear Platform Size Recommendation
5.8.1.1 Clear Platform Size Recommendation The Committee recommends a platform size with a minimum clear width of 32 inches and a minimum clear length of 40 inches. Rationale...
-
5.8.1.1 Clear Platform Size
5.8.1.1 Clear Platform Size. The platform should be a minimum clear width of 32-inches and a minimum clear length of 40-inches.
-
M303.3.3.5 Handrails
Ramps with a rise greater than 6 inches (150 mm) shall provide handrails on each side.
-
M303.3.3.3 Clear Width
The clear width of ramp runs shall be 36 inches (915 mm) minimum.