5.2.1.4 Transfer Surface Size Recommendations for Imaging Equipment (M301) The Committee recommends a transfer surface size on imaging equipment be a minimum of 28 inches long by a minimum...
Search Results "Imaging Equipment"
-
5.2.1.4 Transfer Surface Size Recommendations for Imaging Equipment (M301)
-
5.3 Transfer Sides Recommendations
5.3 Transfer Sides Recommendations
-
5.3.1 Permitted Obstructions to Transfer Sides for M301 and M302
The provisions result in the transfer surface being located at a corner of the equipment and the two transfer sides adjoining at the edges of the equipment....
-
5.3.2 Transfer Sides Recommendations for M301
The provisions result in the transfer surface being located at a corner of the equipment and the two transfer sides adjoining at the edges of the equipment (e.g., foot of an examination...
-
5.3.2.2 Transfer Sides Recommendation for Imaging Equipment
5.3.2.2 Transfer Sides Recommendation for Imaging Equipment On imaging equipment, the transfer surface is oriented along the long dimension of the surface on which patients are in the...
-
5.4 Transfer Supports Recommendations
5.4 Transfer Supports Recommendations
-
5.4.1 Transfer Support Location for M301 and M302
, it is recommended that supports be provided on each side of the transfer surface that is 15 inches deep minimum for patients to maintain position after they have transferred onto the equipment...
-
5.4.1.2 Transfer and Positioning Support Location Recommendations for Imaging Equipment with Transfer Surfaces
5.4.1.2 Transfer and Positioning Support Location Recommendations for Imaging Equipment with Transfer Surfaces For imaging equipment with transfer surfaces, the Committee recommends adding...
-
5.4.2 Transfer Support Length for M301 and M302
5.4.2 Transfer Support Length for M301 and M302 Description: Transfer support length applies to the gripping surface of the support, that portion of the support used to grasp or...
-
5.4.2.2 Transfer and Positioning Support Length Recommendations for Imaging Equipment with Transfer Surfaces
5.4.2.2 Transfer and Positioning Support Length Recommendations for Imaging Equipment with Transfer Surfaces On imaging equipment with transfer surfaces, the transfer surface is oriented...
-
5.4.3 Transfer Support Height for M301 and M302
used by patients in a supine, prone, or side-lying position, and diagnostic equipment used by patients in a seated position....
-
5.4.3.1 Positioning Support Height Recommendation for Imaging Equipment with Transfer Surfaces
5.4.3.1 Positioning Support Height Recommendation for Imaging Equipment with Transfer Surfaces For imaging equipment with transfer surfaces, the Committee recommends a positioning support...
-
5.4.4 Transfer Support Distance from Transfer Surface Recommendations for M301 and M302
NPRM Preamble Discussion: The Access Board is considering whether the following technical criteria would be appropriate for the location and size of transfer supports on diagnostic equipment...
-
5.4.4.2 Transfer Support Distance from Transfer Surface Recommendation for Imaging Equipment
5.4.4.2 Transfer Support Distance from Transfer Surface Recommendation for Imaging Equipment The Committee recommends a maximum distance of 1 1/2 inches from the transfer surface to the...
-
5.4.5 Transfer Support Position Recommendation for M301 and M302
It addresses angled supports or contoured portions of horizontal supports often provided for ergonomics or equipment design considerations....
-
5.4.5.1 Transfer and Positioning Support Position Recommendation for Imaging Equipment
5.4.5.1 Transfer and Positioning Support Position Recommendation for Imaging Equipment On imaging equipment, the transfer surface is oriented along the long dimension of the surface where...
-
5.4.6 Transfer Support Gripping Surface Cross Section and Clearances for M301, M302, and M305.2
The Access Board is considering whether the above cross section dimensions would be appropriate for the gripping surfaces of transfer supports on diagnostic equipment used by patients in...
-
5.4.7 Transfer Support Gripping Surface Clearance Recommendation for M301 and M302
This clearance would allow the fingers and knuckles to clear adjacent equipment parts as they grip the support....
-
5.4.8 Transfer Support Gripping Surface Hazard Recommendation for M301 and M302
5.4.8 Transfer Support Gripping Surface Hazard Recommendation for M301 and M302 Description: Gripping surface hazards are sharp or abrasive elements that prevent an individual...
-
5.4.9 Interruptions along Transfer Support Gripping Recommendation for M301 and M302
5.4.9 Interruptions along Transfer Support Gripping Recommendation for M301 and M302 Description: Transfer supports may contain elements to provide structural support or prevent...
-
5.4.10 Transfer Support Fittings Recommendations for M301 and M302
5.4.10 Transfer Support Fittings Recommendations for M301 and M302 Description: To ensure safe and effective transfers, transfer supports must not rotate within their fittings...
-
5.7 Lift Compatibility Recommendations
5.7 Lift Compatibility Recommendations
-
5.7.1 Lift Compatibility Clearance in Base for M301
maneuver far enough over the equipment to safely lower and raise the patient onto and off examination surfaces....
-
5.7.1.2 Lift Compatibility Clearance in Base for Imaging Equipment (M301)
5.7.1.2 Lift Compatibility Clearance in Base for Imaging Equipment (M301) The Committee recommends the use of alternative means of access by an overhead lift in place of clearances in...