The low transfer position height, Item A of M302.2.1, shall cease to have effect on [Insert date 5 years after date of publication in the Federal Register]. [sic]...
Search Results "Transfer Position"
-
M302.2.2 Sunset
-
5.4.1 Transfer Support Location for M301 and M302
Provisions require transfer supports be provided for use with the transfer sides and be located within reach of the transfer surface and not obstruct transfer onto the surface when in position...
-
5.3.3.1 Transfer Sides Recommendation for Exam Chairs with Fixed Footrests
right transfer....
-
5.4.4 Transfer Support Distance from Transfer Surface Recommendations for M301 and M302
surface and can be effectively used to position during a transfer....
-
M301.2.1 Height
The height of the transfer surface during patient transfer shall be 17 inches (430 mm) minimum and 19 inches (485 mm) maximum measured from the floor to the top of the transfer surface....
-
19 Inch Low Height
This memorandum presents the factors that support a minimum standard of 19 inches as the highest point on the transfer surface in a table or chair’s lowest adjustable position....
-
M301.3.1 Transfer Supports (Section-by-Section Analysis)
The MDE NPRM proposed to require transfer supports to be provided for use with transfer sides on diagnostic equipment used by patients in the supine, prone, or side-lying position, and that...
-
5.2.2.2 Transfer Surface Depth Recommendation for M302
5.2.2.2 Transfer Surface Depth Recommendation for M302 The Committee recommends a minimum depth requirement of 17 inches for equipment used by patients in seated positions....
-
5.1.3 Transfer Surface Low Height Options for M301 and M302
5.1.3 Transfer Surface Low Height Options for M301 and M302 The Committee did not agree on a final low height recommendation....
-
Amusement Ride Seats Designed for Transfer: ADA Standard Section 1002.5
Amusement ride seats designed for transfer shall comply with 1002.5 when positioned for loading and unloading. Advisory 1002.5 Amusement Ride Seats Designed for Transfer....
-
5.2.2.1 Transfer Surface Width Recommendations for M302
5.2.2.1 Transfer Surface Width Recommendations for M302 The Committee recommends 21 inches for the minimum width of the transfer surface of examination chairs and other equipment primarily...
-
Clearances Around Gripping Surfaces
The Access Board is considering whether 1½ inches minimum clearance around the gripping surface would be appropriate for transfer supports on diagnostic equipment used by patients in a supine...
-
M301.2.1 Adjustability
Transfer surfaces shall be adjustable in height measured from the floor to the top of the uncompressed transfer surface and shall provide the following: A....
-
M302.3.1 Transfer Supports
Transfer supports shall be provided for use with the transfer sides required by M302.2.3 and shall comply with M305.2....
-
M302.3.1 Transfer Supports
Transfer supports shall be provided for use with transfer sides required by M302.2.4 and shall comply with M305.2.1.1, M305.2.2.1, and M305.2.3 through M305.2.8....
-
17 Inch Low Height
of grab bars on a person’s ability to transfer....
-
11B-1008.3.1.3 Transfer space
A transfer space complying with Sections 11B-305.2 and 11B-305.3 shall be provided adjacent to the transfer platform....
-
11B-1008.3.1.3 Transfer space
A transfer space complying with Sections 11B-305.2 and 11B-305.3 shall be provided adjacent to the transfer platform....
-
11B-1008.3.2.3 Transfer supports
At least one means of support for transferring shall be provided. Advisory 11B-1008.3.2.3 Transfer supports....
-
11B-1008.3.2.3 Transfer supports
At least one means of support for transferring shall be provided. Advisory 11B-1008.3.2.3 Transfer supports....
-
Transfer Height: ADA Standard Section 1002.6.2
The height of transfer device seats shall be 14 inches (355 mm) minimum and 24 inches (610 mm) maximum measured from the load and unload surface....
-
M301.3.1 Transfer Supports
Transfer surfaces required by M301.2 shall provide transfer supports and shall comply with M305.2....
-
M301.3.1 Transfer Supports
Transfer supports shall be provided for use with the transfer sides required by M301.2.3 and shall comply with M305.2....
-
11B-1009.5.4 Transfer steps
Transfer step height shall be 8 inches (203 mm) maximum....