The MDE NPRM proposed that all diagnostic equipment used by patients in a supine, prone, or side-lying position must comply with the technical requirements of proposed section M301....
Search Results "Patient in Prone Position"
-
M301.1 General (Section-by-Section Analysis)
-
M301.2.1 Adjustability (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface height range for diagnostic equipment used by patients in the supine, prone, or side-lying position of 17 inches minimum and 19 inches maximum....
-
M301.2.2 Sunset Provision (Section-by-Section Analysis)
As discussed in Section IV.B.1.a. (Significant Changes – Transfer Surface Adjustability), this is a new provision that was added to the final rule in conjunction with the new...
-
M301.2.3 Size (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface size for diagnostic equipment used in the supine, prone, or side-lying position of 30 inches wide and 15 inches deep minimum....
-
M301.2.4 Unobstructed Transfer (Section-by-Section Analysis)
The final rule retains the requirement for unobstructed transfer, but has reworded the requirement to specify that each transfer surface must provide two unobstructed sides for the patient...
-
M301.3 Supports (Section-by-Section Analysis)
This is an introductory section. An editorial change was made to this section as a result of the change in M301.3.2, described below, to replace the word “stirrups” with the term...
-
M301.3.1 Transfer Supports (Section-by-Section Analysis)
The MDE NPRM proposed to require transfer supports to be provided for use with transfer sides on diagnostic equipment used by patients in the supine, prone, or side-lying position, and that...
-
M301.3.2 Leg Supports (Section-by-Section Analysis)
In the MDE NPRM, the Access Board proposed to place the requirements for stirrups on diagnostic equipment used by patients in the supine, prone, or side-lying position in M301....
-
M301.3.3 Head and Back Support (Section-by-Section Analysis)
In the MDE NPRM the Access Board proposed to place the requirements for head and back support for diagnostic equipment used by patients in the supine, prone, or side-lying position in M301...
-
M305 Supports (Section-by-Section Analysis)
M305 in the final rule provides the technical requirements for transfer supports, standing supports, leg supports, and head and back supports. Transfer supports are required for...
-
M305.1 General (Section-by-Section Analysis)
This is an introductory section.
-
M305.2 Transfer Supports (Section-by-Section Analysis)
This is an introductory section. As discussed above in Section IV.E.1. (Significant Changes – Transfer Supports), the Access Board strengthened the transfer support requirements and...
-
M305.2.1 Location (Section-by-Section Analysis)
End transfer supports cover diagnostic equipment used by patients in the supine, prone, or side-lying position with end transfer surfaces, M301.2.3.1 in the final rule, and all diagnostic...
-
M305.2.2 Length (Section-by-Section Analysis)
the side of the patient table 12 to 16 inches and is located at a position to accommodate clinical use....
-
M305.2.3 Height (Section-by-Section Analysis)
In the MDE NPRM there was no specific requirement regarding the height of the transfer support, only that it be “within reach” of the patient (proposed M305.2.1)....
-
M305.2.4 Cross Section (Section-by-Section Analysis)
The proposed rule did not provide specific requirements for the cross section of transfer supports. However, in the MDE NPRM preamble, the Access Board noted that it was considering...
-
M305.2.5 Surface Hazards (Section-by-Section Analysis)
were submitted on this issue, but the MDE Advisory Committee voiced concern about surface hazards stating, “gripping surface configurations must provide an effective and safe surface for patients...
-
M305.2.6 Gripping Surfaces (Section-by-Section Analysis)
The MDE Advisory Committee explained that: [t]ransfer supports may contain elements to provide structural support or prevent patient entrapment....
-
M305.2.7 Clearance (Section-by-Section Analysis)
In the MDE NPRM, the Access Board did not provide any specific requirements for clearances around the transfer support. However, in the preamble to the MDE NPRM the Access Board...
-
M305.2.8 Fittings (Section-by-Section Analysis)
Another commenter (manufacturer) pointed out that bed rails, which are common on hospital beds, require a latched position and an unlatched position, which allows them to rotate in their...
-
M305.4 Leg Supports (Section-by-Section Analysis)
The MDE NPRM proposed that where stirrups are provided, they must provide a method to support, position, and secure the patients legs....
-
M305.5 Head and Back Support (Section-by-Section Analysis)
The MDE NPRM proposed to require diagnostic equipment used by patients in the supine, prone, or side-lying position and the seated position that can be adjusted to a reclined position to...
-
Figure M3: Overview of M302 Diagnostic Equipment for Use by a Patient in a Seated Position
-
M301 Diagnostic Equipment Used by Patients in Supine, Prone, or Side-Lying Position