Where diagnostic equipment used by patients in a supine, prone, or side-lying position, and diagnostic equipment used by patients in a seated position can be adjusted to reclined positions...
Search Results "Side Reach"
-
Head and Back Support (M301.3.3 and M302.3.3)
-
5.4.2.2 Transfer and Positioning Support Length Recommendations for Imaging Equipment with Transfer Surfaces
Surfaces On imaging equipment with transfer surfaces, the transfer surface is oriented along the long dimension of the surface on which patients are positioned in the supine, prone, or side-lying...
-
5.2.1.1 Transfer Surface Width Recommendations for M301
5.2.1.1 Transfer Surface Width Recommendations for M301 The Committee recommends a transfer surface of 28 inches minimum width for equipment used by patients in a supine, side-lying,...
-
5.4.2.1 Transfer Support Length for Stretchers
5.4.2.1 Transfer Support Length for Stretchers. Transfer supports on stretchers should be a minimum of 15 inches long.
-
5.4.4 Transfer Support Distance from Transfer Surface Recommendations for M301 and M302
Transfer supports shall be located within reach of the transfer surface and shall not obstruct transfer onto or off the surface when in position....
-
Reach Range
Accessible reach ranges specify a minimum height of 15 inches (308.3) for unobstructed reaches....
-
M301.2.3 Size (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface size for diagnostic equipment used in the supine, prone, or side-lying position of 30 inches wide and 15 inches deep minimum....
-
Doors
DOORS 32” clear width 11B-404.2.3 Maneuvering clearances 11B-404.2.4.1 Level landings each side...
-
Doors
DOORS 32” clear width 11B-404.2.3 Maneuvering clearances 11B-404.2.4.1 Level landings each side...
-
A4.17.3 Size and Arrangement
A standard stall provides a clear space on one side of the water closet to enable persons who use wheelchairs to perform a side or diagonal transfer from the wheelchair to the water closet...
-
M301.1 General
Where diagnostic equipment is used by patients in a supine, prone, or side-lying position, it shall comply with M301....
-
M301.2.3 Size
The size of transfer surfaces shall be measured from center points of their opposing sides....
-
5.2.1.1 Transfer Surface Width
The transfer surface should be 28 inches wide minimum for equipment used by patients in a supine, side-lying, or prone position....
-
5.2.1.2 Transfer Surface Depth
The transfer surface should be 17 inches deep minimum for all equipment used by patients in a supine, side-lying, or prone position with the exception of imaging equipment....
-
M301.1 General
Diagnostic equipment that supports patients in a supine, prone, or side-lying position shall comply with M301. ...
-
M301.2.1 Transfer Surface Height
The table side covers have the same issues discussed for the CT table in Figure 5.3-1....
- Midmark 647 Barrier-Free® Power Podiatry Procedures Chair
-
M305.2.3 Fittings
M305.2.3 Fittings Proposed requirement: Transfer supports shall not rotate within their fittings • Sub-Committee Recommendation: Transfer supports shall not rotate within...
-
Overhead Lifts
The technical criteria do not address overhead lifts that are usually mounted on the ceiling and operate on tracks suspended over the diagnostic equipment because the configuration...
-
Clearance in Base (M301.4.1 and M302.4.1)
Clearance in the base of the equipment allows the legs of a portable floor lift to fit under the base of the equipment. The clearance can be an open area between the supporting posts...
- Landscape Forms Inc. Sort Recycling Units
-
M301.3.1 Transfer Supports (Section-by-Section Analysis)
The MDE NPRM proposed to require transfer supports to be provided for use with transfer sides on diagnostic equipment used by patients in the supine, prone, or side-lying position, and that...
-
M301.3.2 Leg Supports (Section-by-Section Analysis)
In the MDE NPRM, the Access Board proposed to place the requirements for stirrups on diagnostic equipment used by patients in the supine, prone, or side-lying position in M301....