handholds, surface softness, number of transfers needed to go from the initial location to the final destination, use of assist devices, space available for transfer, and obstacles or barriers...
Search Results "Transfer Surface Height"
-
Conclusions
-
M301.2.1 Adjustability
Transfer surfaces shall be adjustable in height measured from the floor to the top of the uncompressed transfer surface and shall provide the following: A....
-
M301.2.1 Height
The height of the transfer surface during patient transfer shall be 17 inches (430 mm) minimum and 19 inches (485 mm) maximum measured from the floor to the top of the transfer surface....
-
M301.2.2 Sunset
The low transfer position height, Item A of M301.2.1, shall cease to have effect on [Insert date 5 years after date of publication in the Federal Register]....
-
M302.2.1 Height
The height of the transfer surface during patient transfer shall be 17 inches (430 mm) minimum and 19 inches (485 mm) maximum measured from the floor to the top of the transfer surface....
-
M302.2.1 Adjustability
Transfer surfaces shall be adjustable in height measured from the floor to the top of the uncompressed transfer surface and shall provide the following: A....
-
M302.2.2 Sunset
The low transfer position height, Item A of M302.2.1, shall cease to have effect on [Insert date 5 years after date of publication in the Federal Register]. [sic]...
-
Transfer Protocols
After each transfer in each protocol, any changes made to device positioning (if any), use of a transfer board, and surface(s) used for the leading hand (e.g. reaching to new surface) and...
-
Results
-
Transfer Protocols
surface independently including a standard therapy mat table which was in the testing room....
-
Highest and Lowest Heights Obtained
The 5th percentile lowest and highest transfers attained were similar to the average WMD seat height (22.1 inch) for all protocols....
-
Effect of Handhelds on Transfer Height
Only a few people used the backrests during the transfers....
-
Implication for Standards and Design
The station used in this second phase permitted greater freedom of wheelchair positioning around the transfer surface compared to the first phase station. ...
-
Transfer Heights
Both of these heights are outside of the standards for heights for most elements where adults would be expected to transfer (between 17 and 19 inches [7]). ...
-
4.1.2 Surface Design of Examination Tables and Chairs
This contouring complicates efforts to measure the distance between the floor and the height of the surface onto which patients transfer for their examinations....
-
4.1.3 Adjustable Height Chairs with and without Footplates
Current chair configuration and anthropometric factors inform the implications of minimum height standards for accessibility of chair transfer surfaces for future designs (Section 5.1.3)...
-
Clear Space Evaluation
moving toward a new surface. ...
-
4.3.2 Transfer Surfaces and Imaging Equipment Functions
4.3.2 Transfer Surfaces and Imaging Equipment Functions The transfer surface (table) of diagnostic imaging equipment serves two purposes: (1) it allows patients to be positioned properly...
-
Height
The Access Board is considering whether 6 inches minimum and 19 inches maximum above the transfer surface would be an appropriate height for transfer supports on diagnostic equipment used...
-
Conclusion
to floor height (22.1 inches) and to a platform that was 27 inches wide....
-
5.1 Transfer Surface Height Recommendations
5.1 Transfer Surface Height Recommendations Description: The transfer height is set above the finished floor surface to permit transfers form mobility devices onto the transfer surface...
-
5.1.1 Transfer Surface Height Adjustability Recommendation for M301 and M302
5.1.1 Transfer Surface Height Adjustability Recommendation for M301 and M302 The Committee recommends adjustable height in small, virtually continuous increments....
-
5.1.2 Transfer Surface High Height Recommendations for M301 and M302
5.1.2 Transfer Surface High Height Recommendations for M301 and M302 The Committee recommends an adjustable height range with an upper height measuring at least 25 inches....
-
5.1.3 Transfer Surface Low Height Options for M301 and M302
5.1.3 Transfer Surface Low Height Options for M301 and M302 The Committee did not agree on a final low height recommendation....