5.2 Transfer Surface Size Recommendations
Search Results "Height Adjustable"
-
5.2 Transfer Surface Size Recommendations
-
Examples of access:
Exam table that lowers to 17-19 inches from the floor Many people find these tables easier to use. There are many ways to help people get access. See the [images]...
-
M301.4.1 Clearance in Base
[See subsections ...]
-
M301.2.2 Size
The transfer surface shall be 30 inches (760 mm) wide minimum and 15 inches (381mm) deep minimum. Advisory M301.2.2 Size. The size requirements in this section apply only to the...
-
M301.2 Transfer Surface
A transfer surface shall be provided and shall comply with M301.2.
-
M301.1 General
Where diagnostic equipment is used by patients in a supine, prone, or side-lying position, it shall comply with M301.
-
4.3.2 Transfer Surfaces and Imaging Equipment Functions
adjustable....
-
Adjustable height protocol
Adjustable height protocol: From initial setup, only the height of the platform was adjusted incrementally: higher (Figure 3) and lower than the subject’s seat. ...
-
Height
The Access Board is considering whether 6 inches minimum and 19 inches maximum above the transfer surface would be an appropriate height for transfer supports on diagnostic equipment used...
-
Height: ADA Standard Section 1009.5.3
The height of the transfer platform shall comply with 1009.4.2....
-
Height: ADA Standard Section 704.4.1
Advisory 704.4.1 Height....
-
602.4 Spout Height
Spout outlets shall be 36 inches (915 mm) maximum above the finish floor or ground.
-
4.15.2* Spout Height
Spouts shall be no higher than 36 in (915 mm), measured from the floor or ground surfaces to the spout outlet (see Fig. 27(a)).
-
4.15.2* SPOUT HEIGHT
Spouts shall be no higher than 36 in (915 mm), measured from the floor or ground surfaces to the spout outlet (see Fig. 27(a)). Figure 27 Drinking Fountains and Water...
-
602.4 Spout Height
Spout outlets shall be 36 inches (915 mm) maximum above the finish floor or ground.
-
Methodology
This study was approved by the University of Pittsburgh's Institutional Review Board and all participants provided informed consent before participating in any test procedures.
-
Adjustable height protocol
Percentage of subjects that were able to attain a transfer under adjustable height protocol Height (in) % Subjects Adjustable Height % Subjects Side...
- Midmark Ritter 223 Barrier-Free® Power Examination Table
-
Height: ADA Standard Section 1009.4.2
The height of the transfer wall shall be 16 inches (405 mm) minimum and 19 inches (485 mm) maximum measured from the deck....
- Midmark Ritter 222 Barrier-Free® Power Examination Table
-
Height: ADA Standard Section 1008.3.1.2
The height of transfer platforms shall be 11 inches (280 mm) minimum and 18 inches (455 mm) maximum measured to the top of the surface from the ground or floor surface....
-
Mounting Height and Location
Section 703.4 covers the mounting height and location of signs with tactile characters....
-
M301.3.2 Leg Supports
Where stirrups are provided, leg supports shall also be provided and shall comply with M305.4.
-
M301.3.2 Leg Supports (Section-by-Section Analysis)
In the MDE NPRM, the Access Board proposed to place the requirements for stirrups on diagnostic equipment used by patients in the supine, prone, or side-lying position in M301. In...