The base of diagnostic equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the...
Search Results "Exam Chair"
Commonly Searched Documents
-
M301.4.2 Clearance Around Base
-
M301.4.1 Clearance in Base (Section-by-Section Analysis)
The MDE NPRM proposed certain clearance requirements beneath the diagnostic equipment to allow sufficient space for the legs of a portable patient lift to fit underneath the...
-
M301.4.2 Clearance Around Base
The base of the equipment shall provide a clearance 6 inches (150 mm) high minimum measured from the floor and 36 inches (915 mm) deep minimum measured from the edge of the...
-
M301.4.2 Clearance Around Base
[See subsections ...]
-
(3) Adjustability: Transfer Surface Intermediate Heights
Adjustable height MDE, such as exam tables, imaging tables and chairs, will make it possible to position the transfer surface near the height of the seat of the mobility device....
-
Four Specific Situations in which a Seating Accommodation Must be Provided
The four situations are as follows: If the passenger uses an aisle chair to access the aircraft and cannot readily transfer over a fixed aisle armrest, you must provide a seat in...
-
§ 382.40a Boarding assistance for large aircraft
§ 382.40a Boarding assistance for large aircraft. (a) Paragraphs (b) and (c) of this section apply to air carriers conducting passenger operations with aircraft having a seating...
-
Boarding and Deplaning Assistance Where Level-Entry Boarding is Unavailable
Boarding and Deplaning Assistance Where Level-Entry Boarding is Unavailable For aircraft with 19 or more seats operating at airports with 10,000 or more annual enplanements where...
-
Emergency Planning
elevators, does it have areas of refuge (fire-rated spaces where people who cannot use stairs can go to register a call for help and wait for responders), or Are there evacuation chairs...
-
Are health care providers required to treat all persons with HIV or AIDS, regardless of whether the treatment being sought is within the provider’s area of expertise?
Are health care providers required to treat all persons with HIV or AIDS, regardless of whether the treatment being sought is within the provider’s area of expertise? No. A health...
-
M305.4 Leg Supports
Leg supports shall provide a method of supporting, positioning, and securing the patient’s legs.
-
M301.2 Transfer Surface
A transfer surface shall be provided and shall comply with M301.2.
-
M301.2.3 Size
The size of the transfer surface shall comply with M301.2.3.1 or M301.2.3.2. The size of transfer surfaces shall be measured from center points of their opposing sides.
-
M301.2.3 Size (Section-by-Section Analysis)
The MDE NPRM proposed a transfer surface size for diagnostic equipment used in the supine, prone, or side-lying position of 30 inches wide and 15 inches deep minimum. (proposed...
-
5.2.1 Transfer Surface Size Recommendations for M301
5.2.1 Transfer Surface Size Recommendations for M301 Description: The transfer surface is the part of the diagnostic equipment onto which patients who use mobility devices or aids...
-
5.2.1.1 Transfer Surface Width
5.2.1.1 Transfer Surface Width. The transfer surface should be 28 inches wide minimum for equipment used by patients in a supine, side-lying, or prone position.
-
5.2 Transfer Surface Size for Equipment Used by Patients in Supine, Prone, or Side-Lying Position
5.2 Transfer Surface Size for Equipment Used by Patients in Supine, Prone, or Side-Lying Position.
-
5.2.1 Transfer Surface Size
5.2.1 Transfer Surface Size
-
5.2.1.2 Transfer Surface Depth
5.2.1.2 Transfer Surface Depth. The transfer surface should be 17 inches deep minimum for all equipment used by patients in a supine, side-lying, or prone position with the exception...
-
5.2 Transfer Surface Size Recommendations
5.2 Transfer Surface Size Recommendations
-
M301.2.2 Size
The transfer surface shall be 30 inches (760 mm) wide minimum and 15 inches (381mm) deep minimum. Advisory M301.2.2 Size. The size requirements in this section apply only to the...
-
M301.2 Transfer Surface
A transfer surface shall be provided and shall comply with M301.2.