Transfer supports shall not rotate within their fittings when in place for transfer.
Search Results "Patient Position"
Commonly Searched Documents
- 
    
  
  M305.2.8 Fittings
  
- 
    
  
  M305.2.7 Clearance
  Clearance between the transfer support gripping surface and adjacent surfaces or obstructions shall be 1 ½ inches (38 mm) minimum. 
- 
    
  
  M305.2.7 Clearance (Section-by-Section Analysis)
  In the MDE NPRM, the Access Board did not provide any specific requirements for clearances around the transfer support. However, in the preamble to the MDE NPRM the Access Board... 
- 
    
  
  M305.2.3 Fittings
  Transfer supports shall not rotate within their fittings. 
- 
    
  
  M305.2.2.2 Side Transfer Supports
  Transfer supports for transfer surfaces complying with M301.2.3.2 shall be 28 inches (710 mm) long minimum and shall be positioned along the width of transfer surfaces.... 
- 
    
  
  M302.3.1 Transfer Supports (Section-by-Section Analysis)
  In the MDE NPRM the Access Board proposed that transfer supports must be provided for use with transfer sides on diagnostic equipment used by patients in the seated position, and that these... 
- 
    
  
  11B-805.5 Patient change areas
  Areas where patients change or are prepared for a procedure shall comply with Section 11B-222.... 
- 
    
  
  5.1.3 Transfer Surface Low Height Options for M301 and M302
  5.1.3 Transfer Surface Low Height Options for M301 and M302 The Committee did not agree on a final low height recommendation. Rationale for final recommendation (more detail... 
- 
    
  
  5.1.3 Transfer Surface Low Height
  5.1.3 Transfer Surface Low Height. No agreement on final low height 
- 
    
  
  5.1.2 Transfer Surface High Height
  5.1.2 Transfer Surface High Height. The transfer surface should have an adjustable height range with an upper height measuring at least 25 inches. 
- 
    
  
  5.1.1 Transfer Surface Height Adjustability
  5.1.1 Transfer Surface Height Adjustability. The transfer surface should have an adjustable height in small virtually continuous increments. 
- 
    
  
  5.2.2.2 Transfer Surface Depth
  The transfer surface should be 17 inches deep minimum for equipment used by patients in a seated position.... 
- 
    
  
  M301.3.3 Head and Back Support
  Where the diagnostic equipment is used in a reclined position, head and back support shall be provided and shall comply with M305.5.... 
- 
    
  
  M305.2.1 Location
  Transfer supports shall comply with M305.2.1.1 or M305.2.1.2 and shall be located 1 ½ inches (38 mm) maximum measured horizontally from the plane defined by the nearest edge of the... 
- 
    
  
  M305.2.4 Cross Section
  Transfer supports shall have a cross section complying with 305.2.4.1 or 305.2.4.2. 
- 
    
  
  M305.2.3 Height
  During use, the tops of transfer support gripping surfaces shall be 6 inches (150 mm) minimum and 19 inches (485 mm) maximum higher than the top of the associated uncompressed... 
- 
    
  
  M305.2.6 Gripping Surface
  Transfer support gripping surfaces shall be continuous along their length and shall not be obstructed along their tops or sides. The bottoms of transfer support gripping surfaces... 
- 
    
  
  M305.2.4 Cross Section (Section-by-Section Analysis)
  The proposed rule did not provide specific requirements for the cross section of transfer supports. However, in the MDE NPRM preamble, the Access Board noted that it was considering... 
- 
    
  
  5.3 Transfer Sides
  5.3 Transfer Sides. 
- 
    
  
  5.3 Transfer Sides Recommendations
  5.3 Transfer Sides Recommendations 
- 
    
  
  5.2.2.1 Transfer Surface Width
  The transfer surface should be 21 inches wide minimum for examination chairs and other equipment used by patients in a seated position.... 
- 
    
  
  C. M303 Diagnostic Equipment Used by Patients Seated in a Wheelchair
  M303 contains the technical requirements for diagnostic equipment used by patients seated in wheelchairs.... 
- 
    
  
  M301.2.1 Adjustability
  A low transfer position at a height of 17 inches (430 mm) minimum and 19 inches (485 mm) maximum; B. A high transfer position at 25 inches (635 mm); and C.... 
- 
    
  
  M305.5 Head and Back Support
  Where the diagnostic equipment is used in a reclined position, head and back support shall be provided.... 