In the MDE NPRM the Access Board proposed that each transfer side provide unobstructed access to the transfer surface, with an exception to permit temporary obstructions as long as...
Search Results "Imaging Bed"
-
M301.2.4 Unobstructed Transfer (Section-by-Section Analysis)
-
M301.4 Lift Compatibility
Diagnostic equipment shall be usable with a patient lift and shall comply with M301.4.1 or M301.4.2.
-
M301.2.3 Transfer Sides
The transfer surface shall be located to provide options to transfer from a mobility device onto one short side (depth) and one long side (width) of the surface. Each transfer side...
-
5.4.3 Transfer Support Height for M301 and M302
Members of the committee also discussed IEC 60601-2-52, which defines specific requirements to prevent entrapment safety hazards on specific types of MDE (typically hospital beds and stretchers...
-
M301 Diagnostic Equipment Used by Patients in Supine, Prone, or Side-Lying Position
[See subsections ...]
-
M301 Diagnostic Equipment Used by Patients in Supine, Prone, or Side-Lying Position
[See subsections ...]
-
M305.2.2 Length
The length of transfer supports shall comply with M305.2.2.1 or M305.2.2.2.
-
M301.2.1 Height
The height of the transfer surface during patient transfer shall be 17 inches (430 mm) minimum and 19 inches (485 mm) maximum measured from the floor to the top of the transfer...
-
d. Unobstructed Transfer
Additionally, a manufacturer, responding to question 9 in the MDE NPRM, explained that all beds, stretchers, and cots have side rails that can be moved to allow unobstructed access for transfer...
-
M305.2.8 Fittings
Transfer supports shall not rotate within their fittings when in place for transfer.
-
M305.2.7 Clearance
Clearance between the transfer support gripping surface and adjacent surfaces or obstructions shall be 1 ½ inches (38 mm) minimum.
-
M305.2.7 Clearance (Section-by-Section Analysis)
In the MDE NPRM, the Access Board did not provide any specific requirements for clearances around the transfer support. However, in the preamble to the MDE NPRM the Access Board...
-
5.4.10 Transfer Support Fittings Recommendations for M301 and M302
5.4.10 Transfer Support Fittings Recommendations for M301 and M302 Description: To ensure safe and effective transfers, transfer supports must not rotate within their fittings...
-
5.4.10 Transfer Support Fittings for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.10 Transfer Support Fittings for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position. The transfer supports shall not rotate when locked in place for patient...
-
M305.2.3 Fittings
Transfer supports shall not rotate within their fittings.
-
11B-807.2.3 Beds
Where beds are provided, clear floor space complying with Section 11B-305 shall be provided on at least one side of the bed....
-
11B-807.2.3 Beds
Where beds are provided, clear floor space complying with Section 11B-305 shall be provided on at least one side of the bed....
-
36 CFR Part 1195 Proposed Accessibility Standards for Medical Diagnostic Equipment NPRM
standards contain minimum technical criteria to ensure that medical diagnostic equipment, including examination tables, examination chairs, weight scales, mammography equipment, and other imaging...
-
11B-805.2.2 Beds
A 36 inch (914 mm) minimum wide clear space shall be provided along the full length of each side of the beds. [2010 ADA Standards] 805.3 Clear Floor or Ground Space....
-
11B-805.2.2 Beds
A 36 inch (914 mm) minimum wide clear space shall be provided along the full length of each side of the beds. [2010 ADA Standards] 805.3 Clear Floor or Ground Space....
-
36 CFR Part 1195 Proposed Accessibility Standards for Medical Diagnostic Equipment NPRM - Preamble
standards contain minimum technical criteria to ensure that medical diagnostic equipment, including examination tables, examination chairs, weight scales, mammography equipment, and other imaging...
-
M305.2.1 Location
Transfer supports shall comply with M305.2.1.1 or M305.2.1.2 and shall be located 1 ½ inches (38 mm) maximum measured horizontally from the plane defined by the nearest edge of the...
-
M305.2.4 Cross Section
Transfer supports shall have a cross section complying with 305.2.4.1 or 305.2.4.2.
-
M305.2.6 Gripping Surface
Transfer support gripping surfaces shall be continuous along their length and shall not be obstructed along their tops or sides. The bottoms of transfer support gripping surfaces...