The height of the top of the gripping surface on horizontal standing supports shall be 34 inches (865 mm) minimum and 38 inches (965 mm) maximum above the standing surface....
Search Results "Guy Support"
-
M305.3.2.1 Horizontal Position
-
M305.3.1.1 Horizontal Position
The length of gripping surfaces on horizontal standing supports shall be 4 inches (100 mm) minimum except for diagnostic equipment containing a wheelchair space which shall comply with M305.3.1.2...
-
5.4.5.1 Transfer and Positioning Support Position Recommendation for Imaging Equipment
5.4.5.1 Transfer and Positioning Support Position Recommendation for Imaging Equipment On imaging equipment, the transfer surface is oriented along the long dimension of the surface where...
-
M305 Supports (Section-by-Section Analysis)
M305 in the final rule provides the technical requirements for transfer supports, standing supports, leg supports, and head and back supports....
-
1193.33(b) (Advisory Guidance)
Paragraph (b) 1. This paragraph requires manufacturers to supply a point of contact for obtaining information about accessibility features of the product and how to obtain...
-
602.1 General
Documentation that supports the use of ICT shall conform to 602. Advisory 602.1 General. ...
-
5.4.2.2 Transfer and Positioning Support Length for Imaging Equipment with Transfer Surfaces
5.4.2.2 Transfer and Positioning Support Length for Imaging Equipment with Transfer Surfaces....
-
M305.3.2 Vertical Position
Where the support is vertical, it shall be 18 inches (455 mm) minimum in length and the bottom end of the support shall be 34 inches (865 mm) high minimum and 37 inches (940 mm) high maximum...
-
5.9 Standing Support for Equipment Used by Patients in a Standing Position
5.9 Standing Support for Equipment Used by Patients in a Standing Position....
-
602.1 General
Documentation that supports the use of ICT shall conform to 602....
-
M301.3.3 Head and Back Support
Where the diagnostic equipment is used in a reclined position, head and back support shall be provided and shall comply with M305.5....
-
5.9 Standing Support Recommendations for M304 Diagnostic Equipment Used by Patients in a Standing Position
5.9 Standing Support Recommendations for M304 Diagnostic Equipment Used by Patients in a Standing Position Description: Many people who ambulate with a disability need support to maintain...
-
5.4.2.2 Transfer and Positioning Support Length Recommendations for Imaging Equipment with Transfer Surfaces
For imaging equipment with transfer surfaces, the Committee recommends a transfer support be 28 inches minimum length and that a positioning support be 12 to 16 inches in length....
-
vii. Ancillary equipment
Ancillary equipment includes items such as positioning straps or cushions; protective padding; adjustable, padded leg supports for gynecological examinations; and additional supports, rails...
-
M305.3.1.3 Vertical Position
The length of the gripping surface on vertical standing supports shall be 18 inches (455 mm) minimum....
-
Medical Diagnostic Equipment Accessibility Standards Sub-Committee Recommendations Weight Scales
Specifically, section M303, patient positions equipment designed to support seated in a wheelchair....
-
M305.3.2.2 Vertical Position
The height of the lowest end of the gripping surface on vertical standing supports shall be 34 inches (865 mm) minimum and 37 inches (940 mm) maximum above the standing surface....
-
"Doing work'' or "performing tasks.'' (Section-by-Section Analysis)
The Department agrees with this statement but cautions that unless the animal is individually trained to do something that qualifies as work or a task, the animal is a pet or support animal...
-
Doing "work'' or "performing tasks.'' (Section-by-Section Analysis)
The Department agrees with this statement but cautions that unless the animal is individually trained to do something that qualifies as work or a task, the animal is a pet or support animal...
-
Transfer Supports Location and Configuration M301.3 and M302.3
Transfer Supports Location and Configuration M301.3 and M302.3 Location of the transfer supports shall be a minimum of 6 inches to a maximum of 19 inches from the top of the transfer...
-
M301.3 Supports (Section-by-Section Analysis)
An editorial change was made to this section as a result of the change in M301.3.2, described below, to replace the word “stirrups” with the term “leg supports.”...
-
4.4.3 Backrest and Armrest
Allowable stresses shall not be exceeded for materials used when a vertical or horizontal force of 250 pounds (1112 N) is applied at any point on the seat, fastener, mounting device or supporting...
-
M301.3.3 Head and Back Support (Section-by-Section Analysis)
In the MDE NPRM the Access Board proposed to place the requirements for head and back support for diagnostic equipment used by patients in the supine, prone, or side-lying position in M301...
-
M305.5 Head and Back Support (Section-by-Section Analysis)
(Section-by-Section Analysis – M302.3.3), the technical requirements for head and back supports from M301 and M302 have been relocated to M305 Supports....