5.2 Transfer Surface Size Recommendations
Search Results "Patient in Prone Position"
-
5.2 Transfer Surface Size Recommendations
-
5.2.1 Transfer Surface Size Recommendations for M301
5.2.1 Transfer Surface Size Recommendations for M301 Description: The transfer surface is the part of the diagnostic equipment onto which patients who use mobility devices or aids transfer...
-
5.2.1.1 Transfer Surface Width Recommendations for M301
or prone position....
-
5.2.1.3 Transfer Surface Size Recommendations for Stretchers (M301)
An important element is the position at which the patient transfers onto the stretcher. The point of entry to the equipment is in the center....
-
5.2.1.4 Transfer Surface Size Recommendations for Imaging Equipment (M301)
The transfer surface must be located so the 28-inch long dimension is parallel to the patient scanning/imaging side of the table....
-
5.3 Transfer Sides Recommendations
5.3 Transfer Sides Recommendations
-
5.3.1 Permitted Obstructions to Transfer Sides for M301 and M302
Obstructions shall not be vertically less than 1 inch nor be horizontally more than 3 inches from the edge of the patient support surface to provide unobstructed access to the transfer surface...
-
5.3.2 Transfer Sides Recommendations for M301
Patients who use mobility devices would have the choice to approach parallel to the deep dimension of the transfer surface, parallel to the wide dimension of the transfer surface, or at...
-
5.3.2.1 Transfer Sides Recommendation for Stretchers
5.3.2.1 Transfer Sides Recommendation for Stretchers On stretchers, the transfer surface is oriented along the long dimension of the surface on which patients are in the supine, prone...
-
5.3.2.2 Transfer Sides Recommendation for Imaging Equipment
supine, prone, or side-lying position (see Figure 5.2.1.4)....
-
5.4 Transfer Supports Recommendations
5.4 Transfer Supports Recommendations
-
5.4.1 Transfer Support Location for M301 and M302
Transfer supports shall be located within reach of the transfer surface and shall not obstruct transfer onto or off the surface when in position....
-
5.4.1.1 Transfer Support Location Recommendation for Stretchers
Rationale for the recommendation Patients access stretchers on either of the long sides rather than on its end (short side)....
-
5.4.1.2 Transfer and Positioning Support Location Recommendations for Imaging Equipment with Transfer Surfaces
Generally, for positioning supports, the table is supporting the full weight of the patient rather than the support....
-
5.4.2 Transfer Support Length for M301 and M302
The purpose of a minimum length is to ensure that there is a minimum gripping surface available to support positioning and repositioning during a transfer....
-
5.4.2.1 Transfer Support Length Recommendation for Stretchers
5.4.2.1 Transfer Support Length Recommendation for Stretchers On Stretchers, the transfer surface is oriented along the long dimension of the surface on which patients are positioned...
-
5.4.2.2 Transfer and Positioning Support Length Recommendations for Imaging Equipment with Transfer Surfaces
along the long dimension of the surface on which patients are positioned in the supine, prone, or side-lying position (see Figure 5.2.1.4)....
-
5.4.3 Transfer Support Height for M301 and M302
The purpose of a minimum height is to ensure that the gripping surface is at an adequate height to facilitate positioning during a transfer....
-
5.4.4 Transfer Support Distance from Transfer Surface Recommendations for M301 and M302
Discussion: The Access Board is considering whether the following technical criteria would be appropriate for the location and size of transfer supports on diagnostic equipment used by patients...
-
5.4.4.1 Transfer Support Distance from Transfer Surface Recommendation for Stretchers
The horizontal distance from transfer surface should be no more than 3 inches from the edge of the patient support surface....
-
5.4.4.2 Transfer Support Distance from Transfer Surface Recommendation for Imaging Equipment
Distance from Transfer Surface Recommendation for Imaging Equipment The Committee recommends a maximum distance of 1 1/2 inches from the transfer surface to the support (either transfer or positioning...
-
5.4.5 Transfer Support Position Recommendation for M301 and M302
5.4.5 Transfer Support Position Recommendation for M301 and M302 Description: Transfer support position refers to the placement of the gripping surface within the minimum and maximum...
-
5.4.5.1 Transfer and Positioning Support Position Recommendation for Imaging Equipment
patients lie in a supine, prone, or side-lying position (see Figure 5.4.2.2)....
-
5.4.6 Transfer Support Gripping Surface Cross Section and Clearances for M301, M302, and M305.2
The Access Board is considering whether the above cross section dimensions would be appropriate for the gripping surfaces of transfer supports on diagnostic equipment used by patients in...