5.2 Transfer Surface Size for Equipment Used by Patients in Supine, Prone, or Side-Lying Position.
Search Results "Exam Table"
-
5.2 Transfer Surface Size for Equipment Used by Patients in Supine, Prone, or Side-Lying Position
-
5.2.1 Transfer Surface Size
5.2.1 Transfer Surface Size
-
5.2.1.2 Transfer Surface Depth
5.2.1.2 Transfer Surface Depth. The transfer surface should be 17 inches deep minimum for all equipment used by patients in a supine, side-lying, or prone position with the exception...
-
5.2 Transfer Surface Size Recommendations
5.2 Transfer Surface Size Recommendations
-
M301.2.2 Size
The transfer surface shall be 30 inches (760 mm) wide minimum and 15 inches (381mm) deep minimum. Advisory M301.2.2 Size. The size requirements in this section apply only to the...
-
M301.2 Transfer Surface
A transfer surface shall be provided and shall comply with M301.2.
-
M305.2.1 Location (Section-by-Section Analysis)
The Committee noted that: because of the size, diversity, and use of diagnostic imaging tables, this support will carry out different functions on different tables… This two-part...
-
11B-226.1 General
to procure or adjust work stations such as desks, laboratory and work benches, fume hoods, reception counters, teller windows, study carrels, commercial kitchen counters, and conference tables...
-
5.4.4 Transfer Support Distance from Transfer Surface Recommendations for M301 and M302
The above technical criteria would likely result in the transfer surface being located at the foot end of examination tables and allow the use of transfer supports similar to the side rails...
-
M305.2.1 Location
Transfer supports shall comply with M305.2.1.1 or M305.2.1.2 and shall be located 1 ½ inches (38 mm) maximum measured horizontally from the plane defined by the nearest edge of the...
-
M305.2.4 Cross Section
Transfer supports shall have a cross section complying with 305.2.4.1 or 305.2.4.2.
-
M305.2.3 Height
During use, the tops of transfer support gripping surfaces shall be 6 inches (150 mm) minimum and 19 inches (485 mm) maximum higher than the top of the associated uncompressed...
-
M305.2.6 Gripping Surface
Transfer support gripping surfaces shall be continuous along their length and shall not be obstructed along their tops or sides. The bottoms of transfer support gripping surfaces...
-
M305.2.6 Gripping Surfaces (Section-by-Section Analysis)
The proposed rule did not provide any specific requirements regarding gripping surfaces on transfer supports. However, in the MDE NPRM preamble the Access Board repeatedly noted that...
-
M305.2.4 Cross Section (Section-by-Section Analysis)
The proposed rule did not provide specific requirements for the cross section of transfer supports. However, in the MDE NPRM preamble, the Access Board noted that it was considering...
-
5.4.9 Interruptions along Transfer Support Gripping Recommendation for M301 and M302
5.4.9 Interruptions along Transfer Support Gripping Recommendation for M301 and M302 Description: Transfer supports may contain elements to provide structural support or prevent...
-
5.4.1 Transfer Support Location for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.1 Transfer Support Location for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position. Transfer supports should be required on both sides of the transfer surface and...
-
5.4.5 Transfer Support Position Recommendation for M301 and M302
5.4.5 Transfer Support Position Recommendation for M301 and M302 Description: Transfer support position refers to the placement of the gripping surface within the minimum and...
-
5.4.2 Transfer Support Length for M301 and M302
5.4.2 Transfer Support Length for M301 and M302 Description: Transfer support length applies to the gripping surface of the support, that portion of the support used to grasp or...
-
5.4.6 Transfer Support Gripping Surface Cross Section for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.6 Transfer Support Gripping Surface Cross Section for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position. The gripping surface cross section should comply with the...
-
5.3 Transfer Sides
5.3 Transfer Sides.
-
5.4.2 Transfer Support Length for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.2 Transfer Support Length for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position. The transfer support should be a minimum length of 15 inches positioned so that...
-
5.3 Transfer Sides Recommendations
5.3 Transfer Sides Recommendations
-
5.4.5 Transfer Support Position for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position
5.4.5 Transfer Support Position for Equipment Used in a Supine, Prone, Side-Lying, or Seated Position. The transfer gripping surface should be located within the minimum and maximum...