5.2.1.3 Transfer Surface Size Recommendations for Stretchers (M301) The Committee recommends a minimum transfer surface size of 28 inches wide by 17 inches deep for stretchers located...
Search Results "Transfer Side"
Commonly Searched Documents
-
5.2.1.3 Transfer Surface Size Recommendations for Stretchers (M301)
-
5.4.4.2 Transfer Support Distance from Transfer Surface Recommendation for Imaging Equipment
5.4.4.2 Transfer Support Distance from Transfer Surface Recommendation for Imaging Equipment The Committee recommends a maximum distance of 1 1/2 inches from the transfer surface to the...
-
5.4.5.1 Transfer and Positioning Support Position Recommendation for Imaging Equipment
patients lie in a supine, prone, or side-lying position (see Figure 5.4.2.2)....
-
M301.2.1, M302.2.1 Transfer Height
M301.2.1, M302.2.1 Transfer Height Proposed requirement: The height of the transfer surface during patient transfer shall be 17 inches (430 mm) minimum and 19 inches (485 mm) maximum...
-
1009.5.1 Transfer Platform
1009.5.1 Transfer Platform. A transfer platform shall be provided at the head of each transfer system....
-
1009.5.1 Transfer Platform
A transfer platform shall be provided at the head of each transfer system....
-
1008.3.1.3 Transfer Space
1008.3.1.3 Transfer Space. A transfer space complying with 305.2 and 305.3 shall be provided adjacent to the transfer platform....
-
1008.3.1.3 Transfer Space
A transfer space complying with 305.2 and 305.3 shall be provided adjacent to the transfer platform....
-
1008.3.1.3 Transfer Space
A transfer space complying with 305.2 and 305.3 shall be provided adjacent to the transfer platform....
-
1009.5.1 Transfer Platform
A transfer platform shall be provided at the head of each transfer system....
-
11B-604.8.1.1.2 Maneuvering space with side-opening door
In a wheelchair accessible compartment with a door located in the side wall or partition, either in-swinging or out-swinging, a minimum 60 inches (1524 mm) wide and 60 inches (1524 mm) deep...
-
Height
The Access Board is considering whether 6 inches minimum and 19 inches maximum above the transfer surface would be an appropriate height for transfer supports on diagnostic equipment used...
-
M301.3.1 Transfer Supports
Transfer supports shall be provided for use with the transfer sides required by M301.2.3 and shall comply with M305.2....
-
Height: ADA Standard Section 1009.5.3
The height of the transfer platform shall comply with 1009.4.2....
-
11B-1009.5.3 Height
The height of the transfer platform shall comply with Section 11B-1009.4.2....
-
Surface: ADA Standard Section 1009.4.4
Surfaces of transfer walls shall not be sharp and shall have rounded edges....
-
Surface: ADA Standard Section 1009.5.5
The surface of the transfer system shall not be sharp and shall have rounded edges....
-
11B-1009.4.4 Surface
Surfaces of transfer walls shall not be sharp and shall have rounded edges....
-
11B-1009.5.5 Surface
The surface of the transfer system shall not be sharp and shall have rounded edges....
-
11B-1009.5.1 Transfer platform
A transfer platform shall be provided at the head of each transfer system....
-
M305.2.1.1 End Transfer Supports
Transfer supports for transfer surfaces complying with M301.2.3.1 and M302.2 shall be located on the short side (length) opposite the transfer side....
- Quadraplegic Transfer
-
11B-1008.3.1.3 Transfer space
A transfer space complying with Sections 11B-305.2 and 11B-305.3 shall be provided adjacent to the transfer platform....
-
11B-1008.3.1.3 Transfer space
A transfer space complying with Sections 11B-305.2 and 11B-305.3 shall be provided adjacent to the transfer platform....