The proposed rule recommended a knee and toe clearance depth for breast platforms of 25 inches. There were no comments received on this requirement. The MDE Advisory Committee...
Search Results "Mammography"
Commonly Searched Documents
-
a. Knee and Toe Clearance
-
5.8.2 Knee and Toe Clearance Under Breast Platform
5.8.2 Knee and Toe Clearance Under Breast Platform. The overall knee and toe space should be increased to a minimum of 28 inches deep, from the initially proposed 25 inch absolute...
-
5.8.2.5 Toe Height
5.8.2.5 Toe Height. The toe height should be 18 inches minimum, measured above the ground when the top of the breast platform is 34 inches above the ground.
-
5.8.2.7 Base Support Allowance
5.8.2.7 Base Support Allowance. Base support is permitted to obstruct the clear floor space if it fits within the allowed base support volume. Base supports can be a maximum of 1½...
-
5.8.2.2 Knee Clearance Depth at 27” Above the Ground
5.8.2.2 Knee Clearance Depth at 27” Above the Ground. The knee clearance should be increased at 27 inches above the ground to 18 inches minimum.
-
5.8.2.2 Knee Clearance Depth at 27” Above the Ground
5.8.2.2 Knee Clearance Depth at 27” Above the Ground The Committee recommends increasing the knee clearance at 27 inches above the ground to 18 inches minimum.
-
5.8.2.4 Unobstructed Floor Space
5.8.2.4 Unobstructed Floor Space. The unobstructed floor space in front of the base support should be a minimum of 17 inches deep.
-
5.8.3.1 Breast Platform Height
5.8.3.1 Breast Platform Height. The height of the top of the breast platform should be decreased to 26 inches above the floor. The upper height range for the breast platform was not...
-
5.8.2.6 Clearance Depth at Toe Height Above the Ground
5.8.2.6 Clearance Depth at Toe Height Above the Ground. The toe clearance depth should be 22 inches minimum at toe height.
-
5.8.3 Breast Platform
5.8.3 Breast Platform
-
5.8.2.3 Overall Knee and Toe Clearance Recommendation
5.8.2.3 Overall Knee and Toe Clearance Recommendation The Committee recommends increasing the overall knee and toe space to 28 inches minimum. Rationale for the...
-
5.8.2.6 Clearance Depth at Toe Height above the ground Recommendation
5.8.2.6 Clearance Depth at Toe Height above the ground Recommendation The Committee recommends 22 inches minimum clearance depth at toe height. Rationale for...
-
5.8.2.3 Overall Knee and Toe Clearance
5.8.2.3 Overall Knee and Toe Clearance. The overall knee and toe space should be increased to a minimum of 28 inches.
-
Figure M6: M303.2.4 Knee and Toe Clearance
[See subsections ...]
-
M305.3.2 Vertical Position
Where the support is vertical, it shall be 18 inches (455 mm) minimum in length and the bottom end of the support shall be 34 inches (865 mm) high minimum and 37 inches (940 mm) high...
-
M303.2.5 Surfaces (Section-by-Section Analysis)
The MDE NPRM proposed to require diagnostic equipment used by patients seated in a wheelchair to provide a wheelchair space with a surface that does not slope more than 1:48 in any...
-
M303.2.5 Surfaces
Wheelchair space surfaces shall not slope more than 1:48 in any direction.
-
M303.2.5 Surfaces
Wheelchair space surfaces shall not slope more than 1:48 in any direction.
-
M303 Diagnostic Equipment Used by Patients Seated in a Wheelchair
[See subsections ...]
-
5.8.2.1 Height of Breast Platform At Time of Measurement
5.8.2.1 Height of Breast Platform At Time of Measurement. All knee and toe clearances should be measured when the top of the breast platform is set at 34 inches above the floor.
-
M303 Diagnostic Equipment Used by Patients Seated in a Wheelchair
[See subsections ...]
-
4.4.2 Mammography Equipment and Scooters
4.4.2 Mammography Equipment and Scooters The Committee considered the interaction of women who use scooters with mammography equipment....
-
4.4.3 Implications of Mammography Equipment Components
4.4.3 Implications of Mammography Equipment Components To determine the accessibility standard recommendations, understanding mammography equipment components is essential....
-
36 CFR Part 1195 Standards for Accessible Medical Diagnostic Equipment - Preamble
equipment (MDE Standards) contain minimum technical criteria to ensure that medical diagnostic equipment, including but not limited to, examination tables, examination chairs, weight scales, mammography...